INSTRUCTIONS FOR USE (IFU)

Size: px
Start display at page:

Download "INSTRUCTIONS FOR USE (IFU)"

Transcription

1 !! INSTRUCTIONS FOR USE (IFU) ME1000 SURGICAL IMPACTOR Rx ONLY English (EN)! 1 English

2 TABLE OF CONTENTS INTRODUCTION... 2 GENERAL PRODUCT INFORMATION... 4 HOW SUPPLIED... 4 DEVICE DESCRIPTION... 4 INTENDED USE... 4 CONTRAINDICATIONS, WARNINGS, PRECAUTIONS... 4 SYMBOLS GLOSSARY... 5 SIGNAL WORDS... 6 GENERAL SAFETY INFORMATION... 9 PRODUCT FEATURES AND SPECIFICATIONS ENVIRONMENTAL CONDITIONS OPERATING INSTRUCTIONS SAFETY INSTRUCTIONS BATTERY PACK ATTACHMENT & REMOVAL ADAPTER ATTACHMENT & REMOVAL IMPACT MODE FORWARD AND REVERSE OPERATION CLEANING SAFETY INSTRUCTIONS POINT OF USE INSTRUCTIONS CONTAINMENT INSTRUCTIONS PREPARATION FOR CLEANING INSTRUCTIONS MANUAL CLEANING INSTRUCTIONS LUBRICATION SAFETY INSTRUCTIONS LUBRICATION INSTRUCTIONS INSPECTION AND FUNCTIONAL TESTING SAFETY INSTRUCTIONS INSPECTION INSTRUCTIONS FUNCTIONAL TESTING INSTRUCTIONS PACKAGING SAFETY INSTRUCTIONS PACKAGING METHODS PACKAGING INSTRUCTIONS STERILIZATION SAFETY INSTRUCTIONS STERILIZATION INSTRUCTIONS STORAGE AND HANDLING SAFETY INSTRUCTIONS STORAGE INSTRUCTIONS TROUBLESHOOTING MAINTENANCE AND REPAIR DISPOSAL AND RECYCLING ORDERING INFORMATION ELECTROMAGNETIC COMPATIBILITY INTRODUCTION! 2 English

3 These Instructions for Use is recommended for safe, effective and compliant operation, processing and care for the ME1000 Surgical Impactor. These Instructions for Use is intended for in-service trainers, biomedical equipment technicians, and central supply/sterile processing technicians. Keep and consult these Instructions for Use during the life of the product. NOTES: Medical Enterprises Distribution is referred to as MED in these Instructions for Use. The ME1000 Surgical Impactor and its accessories are referred to as the ME1000 Impact System in these Instructions for Use.! 3 English

4 GENERAL PRODUCT INFORMATION HOW SUPPLIED The ME1000 Surgical Impactor is provided non-sterile. The Surgical Impactor must be cleaned, lubricated, inspected, functional tested and sterilized prior to first use. Protective covers and films must be completely removed before processing. NOTE: If the Surgical Impactor or any of its parts are damaged or missing, please contact your MED Sales Representative for assistance. DEVICE DESCRIPTION The ME1000 Surgical Impactor is a battery-powered impaction tool which provides power to designated surgical instruments and devices through the ME1000 Adapters. INTENDED USE The ME1000 Surgical Impactor is intended for use in delivering power to surgical instruments to impact hard tissue or bone during surgical procedures, and in the placement and removal of implants, cups, nails, wires, pins and other devices, including acetabular cups and acetabular cup liners. The Surgical Impactor is intended for patient use ONLY after careful consideration of the Instructions for Use. It is designed for use by physicians and trained medical personnel ONLY. Use the ME1000 Surgical Impactor ONLY with the ME1000 Adapters and MB1000 Battery Pack as shown below. BATTERY PACK MB1000 BATTERY PACK ADAPTERS ONLY ME1000 Adapters designated for use with ME1000 Surgical Impactor may be used. For use limitations, instructions and ordering information, reference the ME1000 Adapter Instructions for Use and/or contact your MED Sales Representative or MED for additional support. CONTRAINDICATIONS, WARNINGS, PRECAUTIONS The surgeon must evaluate if the ME1000 Surgical Impactor and its accessories, referred to henceforth as (ME1000 Impact System) are appropriate for the intended procedure, based on the designed capabilities of the ME1000 Impact System with regards to bone density, bone strength, bone size, surgical approach and anatomical situation. Additionally, the manufacturer s contraindications and use considerations for the surgical instrument or device which MED has designated for use with the ME1000 Impact System, must be fully understood and considered prior to use. The MED designated surgical instruments and devices for use with the ME1000 Impact System are identified in the ME1000 Adapter Instructions for Use. Contact your MED Sales Representative or MED for additional support. The warnings and cautions contained within these Instructions for Use relate to the operation, processing and care of the ME1000 Surgical Impactor in addition to its use with the MB1000 Battery Pack and ME1000 Adapters.! 4 English

5 SYMBOLS GLOSSARY The following symbols located on the ME1000 Surgical Impactor and its accessories (ME1000 Impact System) and/or labeling are provided and defined in this section. It is recommended that each are studied and understood for proper interpretation.! 5 English

6 !!!!!!!!!!!!!!!!!!! SYMBOL DEFINITION Caution PPE (Personal Protective Equipment) Consult Instructions for Use Device is classified as type B against electrical shock and leakage current. SYMBOL Battery Charger meets CEC efficiency requirements. DEFINITION Combined Canada/U.S. UL listing mark. Manufacturer Temperature Limitation Combined Canada/U.S. UL classification mark. Direct Current Humidity Limitation W Alternating Current Watts Atmospheric Pressure Limitation V Hz Volt Hertz Date of Manufacture db Ah Decibel Ampere-hour SYMBOL Catalog number DEFINITION Serial Number Wh Watt-hour This device should be disposed of in accordance with environmental protection requirements. Rx ONLY Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner. BFP Back Feed Protection Lock/Unlock Non-sterile Class 2 Battery Charger Classified as a Class 2 power supply SIGNAL WORDS! 6 English

7 The following statements are located on the ME1000 Surgical Impactor and its accessories (ME1000 Impact System) and/or labeling and are intended to explain the levels of risk associated with those products. STATEMENT NOTE MEANING Identifies a safety-related issue. ALWAYS comply with this information to prevent patient and/or healthcare worker injury. Identifies a product reliability issue. ALWAYS comply with this information to prevent product damage. Supplements and/or clarifies procedural information.! Date Printed: REV A 7 English

8 ! Date Printed: REV A 8 English

9 GENERAL SAFETY INFORMATION SAFETY INSTRUCTIONS WARNINGS: Before using the Surgical Impactor, READ and UNDERSTAND all instructions contained in these Instructions for Use. Pay particular attention to safety information. Failure to comply with these Instructions for Use may result in injury. Before using the Surgical Impactor with other MED compatible products, READ and UNDERSTAND the Instructions for Use provided with each product to be used. Pay particular attention to safety information. Failure to comply with those instructions may result in injury. Unless otherwise indicated, ONLY use the recommended packaging as indicated in the PACKAGING section of these Instructions for Use for the storage, sterilization and transportation of the Surgical Impactor. The use of other packaging may lead to extended cleaning times, inadequate sterilization and product damage. The Surgical Impactor must be cleaned, lubricated, inspected, functional tested and sterilized prior to first use and after every subsequent use as indicated in these Instructions for Use. Protective covers and films must be completely removed before reprocessing. DO NOT use or process a ME1000 Surgical Impactor for use if damage, wear, corrosion or missing parts are apparent or the inspection criteria are not met. Contact your MED Sales Representative for disposition and replacement. DO NOT use the Surgical Impactor if it has been dropped and appears damaged or is not functioning properly. Contact your MED Sales Representative for disposition and replacement. DO NOT operate or store the Surgical Impactor where flammable or explosive vapors and/or gasses may be present. ONLY use the Surgical Impactor with the MB1000 Battery Pack and ME1000 Adapters indicated in the INTENDED USE section of these Instructions for Use. Use with other equipment may result in injury and/or product damage. DO NOT modify or disassemble the Surgical Impactor. Modification or disassembly is considered misuse and could result in injury. CALIFORNIA PROPOSITION 65 WARNING: This product may contain chemicals, including lead, known to the State of California to cause cancer, birth defects or other reproductive harm. Wash hands after handling. CAUTIONS: ALWAYS handle the Surgical Impactor with care to prevent product damage. DO NOT drop the Surgical Impactor. DO NOT expose the Surgical Impactor to conditions in excess of the limitations indicated in the ENVIRONMENTAL CONDITIONS section of these Instructions for Use. NOTES: The Surgical Impactor is classified as type BF applied part against electrical shock and leakage current. The Surgical Impactor is suitable for use on patients in accordance with UL For important information regarding electromagnetic compatibility (EMC), reference the ELECTROMAGNETIC COMPATIBLITY section within these Instructions for Use.! 9 English

10 ! PRODUCT FEATURES PRODUCT FEATURES AND SPECIFICATIONS The safe use of the Surgical Impactor requires an understanding of the information in these Instructions for Use as well as knowledge of the procedure to be performed. Before use of this product, familiarize yourself with all operating features and safety instructions. DETENT REGION: Mechanical interface allowing proper connection of the ME1000 Adapters to the ME1000 Surgical Impactor ANVIL: Transmits impact energy from the internal impact mechanism to the Locking Collar. LOCKING COLLAR: Rotary mechanism allowing locked and unlocked connection of Adapters which are properly inserted into the Detent Region. TRIGGER: Spring loaded, momentary switch which initiates impact cycling when fully depressed. BATTERY PACK PORT: The Mechanical interface for the MB1000 Battery Pack. The Battery Pack Port is designed to mate with the Battery Pack Raised Rib to enable the Battery Pack Terminals to electrically connect to the Surgical Impactor. BATTERY PACK LATCH: Secures the MB1000 Battery Pack to the ME1000 Surgical Impactor. The Battery Pack Latch must be in the fully opened position to insert a Battery Pack. The Battery Pack Latch is shown in the fully closed position in the image below. GLAMOR CAP: Covers and protects the Anvil while providing pinch point protection between the Locking Collar and the ME1000 Surgical Impactor housing. PRODUCT SPECIFICATIONS Fig. 1! 10 English

11 Motion... Linear Direction Modes... Forward, Backward Impact frequency Hz Switch Type... Momentary Switch Weight (Surgical Impactor Only) lb. (2.2 kg) Dimensions in. L x 3.5 in. W x 8.5 in. H (36.2 cm L x 8.9 cm W x 21.6 cm H) Degree of protection against electrical shock... BF Degree of protection against water penetration... IPX5 Duty Cycle...5 Seconds On/10 Seconds Off* Operating voltage V DC *Thermal limitations: If the Surgical Impactor stops functioning or only allows single impact mode operation after extended use, allow the MB1000 Battery Pack to cool to room temperature before resuming operation. PRODUCT IDENTIFICATION AND SAFETY INFORMATION The product label is located on the side of the Surgical Impactor and contains product identification and safety information. Reference this label for product name, catalog number and serial number. Important safety information and operation ratings are also included on this label. ENVIRONMENTAL CONDITIONS USE TRANSPORTATION* STORAGE Temperature 30 C 86 F 60 C 140 F for max. 72 h 40 C 104 F 10 C 50 F 29 C 20 F 10 C 50 F Relative Humidity 90% 90% 90% 30% 10% 30% Atmospheric Pressure 1000 hpa 1000 hpa 1000 hpa 811 hpa 811 hpa 811 hpa Altitude ft ft. * Transportation conditions were tested according to ASTM D ! 11 English

12 OPERATING INSTRUCTIONS SAFETY INSTRUCTIONS WARNINGS: The Surgical Impactor must be cleaned, lubricated, inspected, functional tested and sterilized prior to first use and after every subsequent use as indicated in these Instructions for Use. Protective covers and films must be completely removed before reprocessing. Following sterilization, allow the Surgical Impactor to cool for at least 1 hour before use. When handling the Surgical Impactor, ALWAYS use personal protective equipment (PPE) including safety goggles as recommended by the healthcare facility guidelines and cleaning agent SDS. The user should ALWAYS use hearing protection when operating the Surgical Impactor. DO NOT operate or store the Surgical Impactor where flammable or explosive vapors and/or gasses may be present. ONLY use the Surgical Impactor for its intended use. To avoid unintended operation, DO NOT place the Surgical Impactor in a magnetic environment. MED recommends an alternative manual system to be readily available during each use of this product in the unlikely event that technical problems occur. Normally, one fully charged MB1000 Battery Pack has sufficient capacity for an entire operation. As a precaution, it is recommended that a second fully charged and sterilized MB1000 Battery Pack is readily available for use so that it can be quickly exchanged under sterile conditions during the surgical procedure, if necessary. To prevent unintended operation, ONLY attach the MB1000 Battery Pack when ready to use Surgical Impactor. Before use, check for misalignment or binding of moving parts, breakage of parts, or any other condition that may affect the Surgical Impactor s operation. Reference the INSPECTION AND FUNCTIONAL TESTING section of these Instructions for Use. MED recommends using MB1000 Battery Packs that have been charged within the last 30 days. DO NOT use or process a ME1000 Surgical Impactor for use if damage, wear, corrosion or missing parts are apparent or the inspection criteria are not met. Contact your MED Sales Representative for disposition and replacement. DO NOT use the Surgical Impactor if it has been dropped and appears damaged or is not functioning properly. Contact your MED Sales Representative for disposition and replacement. Due to vibration of the Surgical Impactor, DO NOT use for more than 3 hours in a 24-hour period. Avoid unintended contact with Trigger when handling the Surgical Impactor with an attached MB1000 Battery Pack. DO NOT use the Surgical Impactor if the Trigger does not turn it on and off. If the Surgical Impactor cannot be controlled with the Trigger, it is dangerous and must be replaced. Contact your MED Sales Representative for disposition and replacement. If the Surgical Impactor continues to operate after the Trigger is released, remove it from the use area, then with assistance, remove the MB1000 Battery Pack. ONLY use the Surgical Impactor with the MB1000 Battery Pack and ME1000 Adapters indicated in the INTENDED USE section of these Instructions for Use. Use with other equipment may result in injury and/or product damage. Before use, ALWAYS ensure that the Locking Collar and Battery Pack Latch are in the locked and/or fully closed position and secured within the Surgical Impactor. If the ME1000 Adapter has a locking mechanism, ALWAYS ensure it is in the locked position and its attachment is secured within the ME1000 Adapter. Unsecured attachments during use may lead to injury. DO NOT use a ME1000 Adapter if it does not have a secure connection to the ME1000 Surgical Impactor. Proper fit of the Adapter with the ME1000 Surgical Impactor is required for proper function. DO NOT use a MB1000 Battery Pack if it does not have a secure connection to the ME1000 Surgical Impactor. When in use, ensure sterile Surgical Impactor components and accessory attachments do not come in contact with non-sterile surfaces outside of the sterile field. DO NOT place hands near moving components during use. During a procedure or reprocessing, ALWAYS place the Surgical Impactor on its side when not being handled to avoid the risk of dropping the device. ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. To prevent damage to the Surgical Impactor and unexpected interruptions in use, ALWAYS follow the duty cycle when using this product. Reference the PRODUCT FEATURES AND SPECIFICATIONS section of these Instructions for Use. A defective Surgical Impactor may not be reused. Reusing a defective Surgical Impactor may lead to injury. Contact your MED Sales Representative for disposition and replacement. DO NOT modify or disassemble the Surgical Impactor. Modification or disassembly is considered misuse and could result in injury.! 12 English

13 DO NOT allow petroleum-based products, penetrating oils, solvents, lubricants or other chemicals to come in contact with the Surgical Impactor unless otherwise specified. These chemicals can lead to inadequate sterilization and/or damage, weaken or destroy device parts that may result in injury. CAUTIONS: ALWAYS handle the Surgical Impactor with care to prevent product damage. DO NOT drop the Surgical Impactor. ALWAYS follow the recommended operating conditions when using the Surgical Impactor. Reference the ENVIRONMENTAL CONDITIONS section of these Instructions for Use. DO NOT strike the Surgical Impactor with a mallet or any other instrument. DO NOT unnecessarily operate the Surgical Impactor without being in contact with an impact surface. NOTES: The user of the Surgical Impactor is responsible for their proper use during surgery. The Surgical Impactor is intended for patient use ONLY after careful consideration of the Instructions for Use. The Surgical Impactor is designed for use by physicians and trained medical personnel ONLY.! 13 English

14 BATTERY PACK ATTACHMENT & REMOVAL Attaching a MB1000 Battery Pack: See Figures 2-3 WARNING: ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. 3.Lower the Battery Pack Latch into the fully closed position when the MB1000 Battery Pack is not attached to prevent damage to the Battery Pack Latch. NOTE: The Surgical Impactor is operational at the point of attachment with a charged MB1000 Battery Pack. 1.If any ME1000 Adapters are connected to the Surgical Impactor, fully extend the Anvil by pulling the Adapter straight out with one hand while holding the Surgical Impactor with the other. 2.Lift the Battery Pack Latch to the fully open position. 3.Align the MB1000 Battery Pack Raised Ribs with the Grooves in the Battery Pack Port and slide the MB1000 Battery Pack into place. 4.Lower the Battery Pack latch until it engages into the fully closed position. 5.While holding the Surgical Impactor with one hand, pull the MB1000 Battery Pack forward and backward to verify a secure, locked connection. Fig. 2 REMOVE Removing a MB1000 Battery Pack: See Figure 4 1.Lift the Battery Pack Latch to the fully open position. 2.Remove the MB1000 Battery Pack by sliding the MB1000 Battery Pack straight out of the Battery Pack Port. Fig. 3! 14 English

15 ADAPTER ATTACHMENT & REMOVAL Fig. 4 ME1000 Adapters are available to connect designated surgical instruments and devices to the Surgical Impactor. The Instructions for Use provided with each ME1000 Adapter indicates the surgical instruments and devices designated for use. Attaching a ME1000 Adapter: See Figures 5-7 WARNING: ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. 1.Remove the MB1000 Battery Pack from the Surgical Impactor. 2.Rotate the Locking Collar to the unlocked position by aligning the arrow symbol on the Locking Collar with the unlock symbol on the Glamor Cap. 3.Insert the ME1000 Adapter into the Detent Region of the Locking Collar until it is fully seated. 4.Rotate the Locking Collar to the locked position by aligning the arrow symbol on the Locking Collar with the lock symbol on the Glamor Cap. Fig. 5 NOTE: The Locking Collar is designed to allow the ME1000 Adapter to be inserted in 90 increments. 5.While holding the Surgical Impactor with one hand, pull the ME1000 Adapter forward and backward to verify a secure, locked connection. Fig. 6 6.Finish by pulling the ME1000 Adapter so that the Anvil is in the fully extended position.! 15 English Fig. 7

16 Removing a ME1000 Adapter: See Figures 8-9 WARNING: ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. 1.Remove the MB1000 Battery Pack from the Surgical Impactor. Fig. 8 2.Rotate the Locking Collar to the unlocked position by aligning the arrow symbol on the Locking Collar with the Unlock Symbol on the Glamor Cap. 3.Remove the ME1000 Adapter by pulling it straight out of the Locking Collar. Fig. 9! 16 English

17 IMPACT MODE The Surgical Impactor delivers power in single impact mode and continuous impact cycle mode. The impact mode is controlled by depressing the Trigger. The Trigger function is described in further detail below. NOTE: During use, observe the listed duty cycle of 5 seconds of use, followed by a 10 second rest for up to 8 cycles. Once 8 cycles are complete, allow the Surgical Impactor to cool for a minimum of 4 minutes before continued use. Single Impact Cycle Mode: 1.Depress and immediately release the Trigger. 2.The Surgical Impactor will deliver 1 impact. Continuous Impact Cycle Mode: 1.Depress and hold the Trigger. 2.The Surgical Impactor will deliver 1 impact, followed by a millisecond delay. 3.Continuous impact cycling will begin following the delay. FORWARD OPERATION 4.Release the Trigger to end impacting. FORWARD AND REVERSE OPERATION The direction of the Surgical Impactor s linear motion is selected by the operator by pushing forward or pulling backward on the Surgical Impactor. Forward Operation: See Figure 10 1.Pushing forward with the Surgical Impactor while depressing the Trigger delivers forward linear motion. Reverse Operation: See Figure 11 1.Pulling backward on the Surgical Impactor while depressing the Trigger delivers backward linear motion. REVERSE OPERATION Fig. 10! 17 English

18 Fig. 11! 18 English

19 CLEANING SAFETY INSTRUCTIONS WARNINGS: The Surgical Impactor must be cleaned, lubricated, inspected, functional tested and sterilized prior to first use and after every subsequent use as indicated in these Instructions for Use. Protective covers and films must be completely removed before reprocessing. Failure to properly clean instruments prior to sterilization may lead to inadequate sterilization. Surgical instruments are used with or on patients who may harbor both recognized and unrecognized infections. To prevent the spread of infection, all reusable instruments must be thoroughly cleaned and sterilized prior to initial use and after each patient use. Instruments may have sharp edges or features. Users and reprocessors must be cautious when handling instruments. ONLY the manual cleaning method as indicated in these Instructions for Use is recommended for cleaning the Surgical Impactor. Other cleaning methods have not been validated and may lead to inadequate cleaning, sterilization and product damage. ONLY individuals trained and experienced in the processing of reusable medical devices should clean the Surgical Impactor. Brushes and other cleaning tools should either be single-use disposable items or, if reusable, decontaminated at least daily. Contaminants could otherwise be transferred from one device to the next during cleaning and pose health risks to healthcare workers. MED recommends the Surgical Impactor to be cleaned using fresh detergent solution. ALWAYS make sure the detergent solution is completely rinsed from the Surgical Impactor before drying the device. DO NOT allow petroleum-based products, penetrating oils, solvents, lubricants or other chemicals to come in contact with the Surgical Impactor unless otherwise specified. These chemicals can lead to inadequate sterilization and/or damage, weaken or destroy device parts that may result in injury. DO NOT use or process a ME1000 Surgical Impactor for use if damage, wear, corrosion or missing parts are apparent or the inspection criteria are not met. Contact your MED Sales Representative for disposition and replacement. ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. When handling the Surgical Impactor, ALWAYS use personal protective equipment (PPE) including safety goggles as recommended by the healthcare facility guidelines and cleaning agent SDS. CAUTIONS: To avoid physical damage with other equipment, ONLY transport the Surgical Impactor in its recommended packaging. Reference the PACKAGING section of these Instructions for Use. DO NOT allow soil to dry on the Surgical Impactor. Maintain moisture on the device as indicated in these Instructions for Use to prevent drying and facilitate cleaning. DO NOT use pressurized water as this may cause damage to the Surgical Impactor. DO NOT use automated or ultrasonic equipment to clean the Surgical Impactor. DO NOT immerse the Surgical Impactor or allow liquid to soak into its terminals or openings. During cleaning, the product may be placed under running water to thoroughly rinse all surfaces. Detergents with a ph of are recommended. Higher or lower ph values can cause damage to the Surgical Impactor materials. ALWAYS prepare the detergent solution per the detergent manufacturer s Instructions for Use. ALWAYS handle the Surgical Impactor with care to prevent product damage. DO NOT drop the Surgical Impactor. During a procedure or reprocessing, ALWAYS place the Surgical Impactor on its side when not being handled to avoid the risk of dropping the device. DO NOT bend the Surgical Impactor terminals during cleaning. DO NOT use sharp, hard, abrasive, pointed or metal cleaning aids (e.g. pipe cleaning, wire brushes, etc.) to clean the Surgical Impactor.! 19 English NOTES: The cleaning and sterilization instructions provided in these Instructions for Use were validated as being capable of preparing this product for use. It is the

20 responsibility of the health care facility to ensure that reprocessing is performed using the appropriate equipment and materials, and that designated personnel have been adequately trained in order to achieve the proper result. ALWAYS verify the cleaning and sterilization processes using your actual equipment and designated personnel to ensure proper results are achieved. DO NOT use tools to remove ME1000 Surgical Impactor accessory attachments or to operate ME1000 Impact System movable part (e.g. locks, latches, etc.) The cleaning instructions in these Instructions for Use apply ONLY to the ME1000 Surgical Impactor. Repeated processing according to these Instructions for Use has minimal effect on and should not compromise the performance of the ME1000 Surgical Impactor. End of life is normally determined by wear and damage due to use. MED used the following cleaning agents during validation of the manual cleaning instructions. These cleaning agents are not listed in preference to other available cleaning agents which may perform satisfactorily. Type: Neutral ph Enzymatic Detergent Product Name: EmPower Dual Enzymatic Detergent Product Catalog Number: Detergents used on the ME1000 Surgical Impactor will be in contact with the following product materials: Stainless steel, Aluminum, Gold, Plastic, Rubber, and Silver/Tin! 20 English

21 POINT OF USE INSTRUCTIONS Recommended Point of Use Supplies for Surgical Impactor: Soft lint-free cloth Tap water Point of Use Steps for Surgical Impactor: See Figures WARNING: ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. NOTE: Reference the OPERATING INSTRUCTIONS section of these Instructions for Use for properly removing the MB1000 Battery Pack and ME1000 Adapters from the Surgical Impactor. 1.Remove the MB1000 Battery Pack from the Surgical Impactor. Fig Remove the ME1000 Adapter from the Surgical Impactor. Fig Position all Surgical Impactor movable parts for cleaning access. Raise the Battery Pack Latch to the fully open position. Fully extend the Anvil. Place the Locking Collar in the locked Position to expose the Lubrication Openings. Fig. 14! 21 English

22 4.Wipe the surfaces of the Surgical Impactor using a clean soft lint-free cloth dampened with clean tap water to remove gross soil and debris. CONTAINMENT INSTRUCTIONS Fig. 15 Recommended Containment Supplies & Devices for Surgical Impactor: Soft lint-free towels Tap water Case Medical SteriTite Full Size Sealed Container w/ Lid w/etched MED Logo Case Medical MediTray Full Size Insert Basket Preconfigured for MED Containment Steps for Surgical Impactor: See Figures Load the Surgical Impactor into its designated place in the MediTray Insert Basket. Place the Anvil in the fully extended position and the Battery Pack Latch in the fully open position. Rotate the Locking Collar to the locked Position to expose the Lubrication Openings. Position the Surgical Impactor in the MediTray Insert Basket with the catalog number facing up. Fig Place the loaded MediTray Insert Basket into the SteriTite Container. Fig. 17! 22 English

23 3.Cover the Surgical Impactor and ME1000 Broach Adapters with clean soft lint-free towels dampened with clean tap water to prevent drying of soil and debris on the device prior to cleaning. Fig Place the SteriTite Container Lid over the SteriTite Container and latch securely for transport. 5.It is recommended that the Surgical Impactor is reprocessed as soon as is reasonably practical following each use. Fig. 18 PREPARATION FOR CLEANING INSTRUCTIONS! 23 English

24 Preparation for Cleaning Steps for Surgical Impactor: See Figures WARNING: ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. NOTES: Reference the OPERATING INSTRUCTIONS section of these Instructions for Use for properly removing the MB1000 Battery Pack and ME1000 Adapters from the Surgical Impactor and for further instruction on Battery Pack Latch and Locking Collar operation. Reference the SteriTite Instructions for Use for properly reprocessing the SteriTite Container and MediTray Insert Basket. 1.Remove the Lid from the SteriTite Container. 2.Remove the Surgical Impactor from the MediTray Insert Basket. Fig Ensure all attachments are removed from the Surgical Impactor. If the MB1000 Battery Pack is connected to the Surgical Impactor, remove the MB1000 Battery Pack first before removing any other attachments. Remove the ME1000 Adapter from the Surgical Impactor if attached. Fig. 21 Fig. 22! 24 English

25 4.Position all Surgical Impactor movable parts for cleaning access. Raise the Battery Pack Latch to the fully open position. Fully extend the Anvil. Rotate the Locking Collar to the locked position to expose the Lubrication Openings Fig. 23 MANUAL CLEANING INSTRUCTIONS Recommended Manual Cleaning Supplies for Surgical Impactor: ph Neutral Enzymatic Detergent Soft lint-free cloth Various sized soft bristle nylon brushes Syringe Tap Water De-ionized (DI) water Filtered pressurized compressed air (20 psi) Fig. 24 Manual Cleaning Steps for Surgical Impactor: See Figures CAUTION: To prevent extended drain times and product damage, DO NOT hold the Battery Pack Port under running water while rinsing, position at an angle to prevent water from entering the Surgical Impactor. 1.Rinse gross soil from all Surgical Impactor surfaces, including Battery Pack Port Terminals, for at least three (3) minutes under cool running tap water. Manipulate the Trigger, Locking Collar and Battery Pack Latch under running water to rinse hidden surfaces. Use a clean soft lint-free cloth, soft bristle nylon brush and/or syringe filled with tap water to assist in removing gross soil from device. Fig Prepare a fresh batch of ph neutral enzymatic detergent solution according to the detergent manufacturer s written instructions.! 25 English

26 3.Using a soft bristled nylon brush wetted with detergent solution, remove any visible soil from the Surgical Impactor by scrubbing all device surfaces with detergent solution for at least five (5) minutes. Manipulate the Trigger, Locking Collar and Battery Pack Latch to scrub hidden surfaces. Scrub all surfaces behind the Locking Collar and inside the Battery Pack Port, including Battery Pack Port Terminals. Scrub the inside of the Locking Collar Opening using an appropriate brush, ensuring the inner walls and the full depth of the Opening are reached. Place the Locking Collar in the Locked Position to access the Lubrication Openings between the Locking Collar Teeth. Flush the inside of each Lubrication Opening using a syringe filled with detergent solution for a minimum of three (3) times. Use a syringe filled with detergent solution to flush any other areas they may be shielded from brushing, ensuring detergent solution contacts all device surfaces. Fig. 26 Fig. 27 Fig. 28! 26 English

27 CAUTION: To prevent extended drain times and product damage, DO NOT hold the Battery Pack Port under running water while rinsing, position at an angle to prevent water from entering the Surgical Impactor. Fig. 30 Fig Rinse all Surgical Impactor surfaces thoroughly under lukewarm running tap water for at least two (2) minutes to remove all traces of detergent solution. Manipulate the Trigger, Locking Collar and Battery Pack Latch under running water to rinse hidden surfaces. Place the Locking Collar in the locked position to access the Lubrication Openings between the Locking Collar Teeth. Flush the inside of each Lubrication Opening using a syringe filled with tap water for a minimum of three (3) times. Use a syringe filled with tap water to flush any other hard-to-reach areas.! 27 English

28 5.Visually inspect all Surgical Impactor surfaces for any remaining soil. Manipulate the Trigger, Locking Collar and Battery Pack Latch to inspect hidden surfaces. Inspect all surfaces behind the Locking Collar and inside the Battery Pack Port. Pay close attention to crevices, grooves, and any hard-to-reach areas. If soil is observed, repeat steps 1-5 using a fresh batch of detergent solution. Fig. 31 CAUTION: To prevent extended drain times and product damage, DO NOT hold the Battery Pack Port under running water while rinsing, position at an angle to prevent water from entering the Surgical Impactor. NOTE: Deionized (DI) water is recommended for the final rinse of the Surgical Impactor. Mineral residues from hard water can stain the product and/or prevent effective cleaning. Fig. 32 DE-IONIZED (DI) WATER RINSE 6.Rinse all Surgical Impactor surfaces thoroughly under de-ionized (DI) water for at least two (2) minutes. Manipulate the Trigger, Battery Pack Latch and Locking Collar under running water to rinse hidden surfaces. Place the Locking Collar in the locked position to access the Lubrication Openings between the Locking Collar Teeth. Flush the inside of each Lubrication Opening using a syringe filled with de-ionized (DI) water for a minimum of three (3) times. Use a syringe filled with de-ionized (DI) water to flush any other hard-to-reach areas.! 28 English Fig. 33 X

29 7.With the Locking Collar in the in the locked position, the Anvil Shaft fully extended and the Battery Pack Latch open, position the Surgical Impactor at a decline on a clean absorbent lintfree cloth with the Locking Collar pointed down. Allow the Surgical Impactor to drain for at least two (2) minutes. Fig Dry all Surgical Impactor surfaces using a clean soft lint-free cloth or filtered pressurized compressed air. Manipulate the Trigger, Battery Pack Latch and Locking Collar to dry hidden surfaces. Pay close attention to crevices, grooves, and openings where moisture can accumulate. Fig. 34! 29 English

30 LUBRICATION Fig. 36! 30 English

31 The Surgical Impactor must be cleaned, lubricated, inspected, functional tested and sterilized prior to first use and after every subsequent use as indicated in these Instructions for Use. Protective covers and films must be completely removed before reprocessing. When handling the Surgical Impactor, ALWAYS use personal protective equipment (PPE) including safety goggles as recommended by the healthcare facility guidelines and cleaning agent SDS. ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. DO NOT modify or disassemble the Surgical Impactor. Modification or disassembly is considered misuse and could result in injury. ONLY use STERILIT I DRIP LUBRICATOR (REF. JG598) to lubricate the ME1000 Surgical Impactor following cleaning as indicated in these Instructions for Use. DO NOT allow petroleum-based products, penetrating oils, solvents, lubricants or other chemicals to come in contact with the Surgical Impactor unless otherwise specified. These chemicals can lead to inadequate sterilization and/or damage, weaken or destroy device parts that may result in injury. After each cleaning and prior to sterilization, ALWAYS perform the lubrication instructions as indicted in these Instructions for Use. Failing to follow these instructions will lead to damage and malfunction, increasing the risk of injury. DO NOT use or process a ME1000 Surgical Impactor for use if damage, wear, corrosion or missing parts are apparent or the inspection criteria are not met. Contact your MED Sales Representative for disposition and replacement. ONLY individuals trained and experienced in the maintenance of reusable medical devices should lubricate the Surgical Impactor. CAUTIONS: ALWAYS handle the Surgical Impactor with care to prevent product damage. DO NOT drop the Surgical Impactor. During a procedure or reprocessing, ALWAYS place the Surgical Impactor on its side when not being handled to avoid the risk of dropping the device. DO NOT strike the Surgical Impactor with a mallet or any other instrument.! 31 English

32 LUBRICATION INSTRUCTIONS Lubricate the Surgical Impactor following the completion of every cleaning and prior to sterilization using STERILIT I Drip Lubricator. Follow the lubricant manufacturer s Instructions for Use (i.e. concentration/dilution and shelf life). Lubrication steps for individual Surgical Impactor components are detailed in the instructions that follow. Reference the SURGICAL IMPACTOR COMPONENT LUBRICATION SITES provided in these instructions for use for assistance in locating each component indicated in the lubrication instructions. See Figures Recommended Lubrication Supplies for Surgical Impactor: STERILIT I DRIP LUBRICATOR REF: JG598 Soft lint-free cloth Lint-free absorbent cloth WARNING: ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. NOTES: A droplet containing a numerical value is provided to show the recommended number of lubrication droplets for each component. See Figure 37 Remove all attachments from the Surgical Impactor prior to lubrication. Fig. 37 Anvil: See Figure 38 1.Fully extend the Anvil by pulling the Locking Collar straight out with one hand while holding the Surgical Impactor with the other hand. If the Anvil is difficult to extend, attach a ME1000 Adapter to the Surgical Impactor to extend the Anvil as indicated in the Anvil OPERATING INSTRUCTIONS section of these Instructions for Use. 2.Put 4 drops of lubricant evenly spaced around the base of the Anvil Shaft. 3.Retract and extend the Anvil Shaft several times to spread lubricant by pushing the Locking Collar inward and pulling it outward. Fig Wipe off excess lubricant from the Glamor Cap and Locking Collar using a clean soft lint-free cloth. Avoid wiping intended lubrication areas. Locking Collar Lubrication Openings: See Figure 39! 32 English

33 1.Place the Locking Collar in the locked position to access Lubrication Openings between Locking Collar Teeth. 2.Put 1 drop of lubricant in each Lubrication Opening. Rotate the Locking Collar to the locked and unlocked position several times to spread lubricant. Locking Collar Lubrication Openings Fig. 39 Detent Region: See Figures Put 2 drops of lubricant on the inner wall inside the Detent Region. 2.Spread the lubricant evenly along the surface using the lubricant applicator tip in a circular action. Detent Region Fig Drain any excess lubricant from the Locking Collar by holding the Surgical Impactor vertically with the Locking Collar pointed down on a clean absorbent cloth for 2 minutes. 4.Wipe off excess lubricant from the Locking Collar using a clean soft lint-free cloth. Avoid wiping intended lubrication areas. Fig. 41! 33 English

34 Trigger Shaft: See Figure 42 1.Position the Surgical Impactor horizontally with the product label side up. 2.Put 1 drop of lubricant on the Trigger Shaft. 3.Depress and release the Trigger several times to spread lubricant. 4.Wipe off excess lubricant from the Trigger housing using a clean soft lint-free cloth. 5.Avoid wiping intended lubrication areas. Trigger Shaft Fig. 42 Battery Pack Latch Hinges: See Figure 43 1.Place the Battery Pack Latch in the locked positon. 2.Put 1 drop of lubricant on each Battery Pack Latch Hinge Pin. 3.Raise and lower the Battery Pack Latch several times to spread lubricant. 4.Wipe off excess lubricant from the Battery Pack Latch using a clean soft lint-free cloth. Avoid wiping intended lubrication areas. Battery Pack Latch Hinges Fig. 43! 34 English

35 Surgical Impactor Component Lubrication Sites: Locking Collar Lubrication Openings Detent Region Battery Pack Latch Hinges Anvil Trigger Shaft Fig. 44! 35 English

36 SAFETY INSTRUCTIONS INSPECTION AND FUNCTIONAL TESTING WARNINGS: The Surgical Impactor must be cleaned, lubricated, inspected, functional tested and sterilized prior to first use and after every subsequent use as indicated in these Instructions for Use. Protective covers and films must be completely removed before reprocessing. When handling the Surgical Impactor, ALWAYS use personal protective equipment (PPE) including safety goggles as recommended by the healthcare facility guidelines and cleaning agent SDS. The user should ALWAYS use hearing protection when operating the Surgical Impactor. DO NOT operate or store the Surgical Impactor where flammable or explosive vapors and/or gasses may be present. ONLY use the Surgical Impactor for its intended use. To prevent unintended operation, ONLY attach the MB1000 Battery Pack when ready to use Surgical Impactor. To avoid unintended operation, DO NOT place the Surgical Impactor in a magnetic environment. Before use, check for misalignment or binding of moving parts, breakage of parts, or any other condition that may affect the Surgical Impactor s operation. Reference the INSPECTION AND FUNCTIONAL TESTING section of these Instructions for Use. DO NOT use or process a ME1000 Surgical Impactor for use if damage, wear, corrosion or missing parts are apparent or the inspection criteria are not met. Contact your MED Sales Representative for disposition and replacement. DO NOT use the Surgical Impactor if it has been dropped and appears damaged or is not functioning properly. Contact your MED Sales Representative for disposition and replacement. Avoid unintended contact with Trigger when handling the Surgical Impactor with an attached MB1000 Battery Pack. DO NOT use the Surgical Impactor if the Trigger does not turn it on and off. If the Surgical Impactor cannot be controlled with the Trigger, it is dangerous and must be replaced. Contact your MED Sales Representative for disposition and replacement. If the Surgical Impactor continues to operate after the Trigger is released, remove it from the use area, then with assistance, remove the MB1000 Battery Pack. ONLY use the Surgical Impactor with the MB1000 Battery Pack and ME1000 Adapters indicated in the INTENDED USE section of these Instructions for Use. Use with other equipment may result in injury and/or product damage. Due to vibration of the Surgical Impactor, DO NOT use for more than 3 hours in a 24-hour period. Before use, ALWAYS ensure that the Locking Collar and Battery Pack Latch are in the locked and/or fully closed position and secured within the Surgical Impactor. If the ME1000 Adapter has a locking mechanism, ALWAYS ensure it is in the locked position and its attachment is secured within the ME1000 Adapter. Unsecured attachments during use may lead to injury. DO NOT use a ME1000 Adapter if it does not have a secure connection to the ME1000 Surgical Impactor. Proper fit of the Adapter with the ME1000 Surgical Impactor is required for proper function. DO NOT use a MB1000 Battery Pack if it does not have a secure connection to the ME1000 Surgical Impactor. DO NOT place hands near moving components during use. ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. To prevent damage to the Surgical Impactor and unexpected interruptions in use, ALWAYS follow the duty cycle when using this product. Reference the PRODUCT FEATURES AND SPECIFICATIONS section of these Instructions for Use. A defective Surgical Impactor may not be reused. Reusing a defective Surgical Impactor may lead to injury. Contact your MED Sales Representative for disposition and replacement. DO NOT modify or disassemble the Surgical Impactor. Modification or disassembly is considered misuse and could result in injury. DO NOT allow petroleum-based products, penetrating oils, solvents, lubricants or other chemicals to come in contact with the Surgical Impactor unless otherwise specified. These chemicals can lead to inadequate sterilization and/or damage, weaken or destroy device parts that may result in injury. Following lubrication and prior to sterilization, ALWAYS perform the inspection and functional testing instructions as indicated in these Instructions for Use. Failing to follow these instructions will lead to damage and malfunction, increasing the risk of injury. ONLY individuals trained and experienced in the maintenance of reusable medical devices should inspect and test the Surgical Impactor. CAUTIONS: ALWAYS handle the Surgical Impactor with care to prevent product damage. DO NOT drop the Surgical Impactor. During a procedure or reprocessing, ALWAYS place the Surgical Impactor on its side when not being handled to avoid the risk of dropping the device.! 36 English

37 DO NOT strike the Surgical Impactor with a mallet or any other instrument. DO NOT unnecessarily operate the Surgical Impactor without being in contact with an impact surface. ALWAYS follow the recommended operating conditions when using the Surgical Impactor. Reference the ENVIRONMENTAL CONDITIONS section of these Instructions for Use. NOTE: The routine inspection and functional testing as indicated in the following instructions, are the best methods for determining the life span of the Surgical Impactor.! 37 English

38 INSPECTION INSTRUCTIONS Inspection Steps for Surgical Impactor: WARNING: ALWAYS remove the MB1000 Battery Pack from the Surgical Impactor before attaching or removing ME1000 Adapters (including surgical instruments and devices designated for use with the Adapters), processing, lubricating and/or performing inspection. Removing the MB1000 Battery Pack will prevent unexpected operation that could cause injury. 1.Visually inspect cleaned Surgical Impactor for damage and wear. Reference the following bulleted examples: Damage: (e.g. cracked, dented or separated housing, physical imperfections that can cause damage or injury, etc.) Wear: (e.g. corrosion, removed or illegible part numbers, symbols and markings, etc.) 2.If damage or wear is detected, contact your MED Sales Representative for disposition and replacement. DO NOT sterilize or allow a ME1000 Surgical Impactor with damage or wear to be reused. 3.If no damage or wear is detected, immediately proceed to the following functional testing instructions. FUNCTIONAL TESTING INSTRUCTIONS Insertion / Removal of Attachments and Movable Parts: See Figures NOTE: Reference the OPERATING INSTRUCTIONS section of these Instructions for Use for insertion/ Removal of attachments and Surgical Impactor Operation assistance while performing the following functional tests. Verify the following criteria while performing the functional tests indicated below. Locking Collar and Battery Pack Latch movement is smooth and without blockage. Trigger depression and release is smooth and without blockage. Insertion and removal of ME1000 Adapter and MB1000 Battery Pack is smooth and without blockage.! 38 English ME1000 Adapter Attachment: See Figures 45-48

39 1.Remove the MB1000 Battery Pack from the Surgical Impactor if attached. Fig Insert a ME1000 Adapter into the Locking Collar and rotate the Locking Collar to the Locked position. Fig While holding the Surgical Impactor with one hand, pull the ME1000 Adapter forward and backward to verify a secure, locked connection. Fig Rotate the Locking Collar to the unlocked position and remove the ME1000 Adapter. MB1000 Battery Pack Attachment and Surgical Impactor Operation: See Figures Fig Remove ME1000 Adapter from Surgical Impactor if attached.! 39 English

40 2.Insert a fully charged MB1000 Battery Pack into Battery Pack Port and lower the Battery Pack Latch to the fully closed position. Fig. 49 Fig While holding the Surgical Impactor with one hand, pull the MB1000 Battery Pack forward and backward to verify a secure, locked connection. Fig. 51! 40 English

41 4. Aim the Surgical Impactor towards the floor, taking care to avoid aiming the Surgical Impactor at anyone. 5.Depress the trigger for 1-2 seconds to operate the Surgical Impactor. DO NOT depress for a longer period since the Surgical Impactor is designed to operate against an impact surface. 6.If the Surgical Impactor does not perform one or more impact cycles, reference the TROUBLESHOOTING section of these Instructions for Use. If the Surgical Impactor continues to not operate properly contact your MED Sales Representative for disposition and replacement. Fig. 52 Fig Remove the MB1000 Battery Pack from the Surgical Impactor after functional testing is complete. 8.Remove the ME1000 Adapter from the Surgical Impactor after the MB1000 Battery Pack has been removed.! 41 English

42 ! 42 English Fig. 54

43 SAFETY INSTRUCTIONS WARNINGS: When handling the Surgical Impactor, ALWAYS use personal protective equipment (PPE) including safety goggles as recommended by the healthcare facility guidelines and cleaning agent SDS. Unless otherwise indicated, ONLY use the recommended packaging as indicated in the PACKAGING section of these Instructions for Use for the storage, sterilization and transportation of the Surgical Impactor. The use of other packaging may lead to extended cleaning times, inadequate sterilization and product damage. ALWAYS make sure the Surgical Impactor has been properly cleaned, lubricated, inspected, functional tested and is completely dry prior to sterile packaging. DO NOT use or process a ME1000 Surgical Impactor for use if damage, wear, corrosion or missing parts are apparent or the inspection criteria are not met. Contact your MED Sales Representative for disposition and replacement. DO NOT use a Lid on the MediTray Insert Basket for packaging the ME1000 Surgical Impactor. The use of a MediTray Lid with the ME1000 Surgical Impactor is not recommended and may lead to inadequate sterilization of the device. ONLY Trays and Containers manufactured for or Distributed by MED should contain MED products as indicated by MED. Use of other containment devices are not recommended and may lead to extended cleaning times, inadequate sterilization and product damage. For Preconfigured Tray, Basket and Container layouts, areas designated for specific devices shall ONLY contain those devices and in the configuration specified by MED. DO NOT add devices to a preconfigured Tray, Basket or Container in excess of its defined device configuration. Overcrowding and/or incorrect loading can result in product damage and impact the effectiveness of sterilization. ONLY load the devices indicated in the TABLE 1 PACKAGING section of these Instructions for Use into the MediTray Insert Basket Preconfigured for MED. CAUTION: ALWAYS handle the Surgical Impactor with care to prevent product damage. DO NOT drop the Surgical Impactor. During a procedure or reprocessing, ALWAYS place the Surgical Impactor on its side when not being handled to avoid the risk of dropping the device. NOTES: The validated sterilization instructions provided in these Instructions for Use are ONLY applicable to the product configurations and packaging methods indicated in the PACKAGING section of these Instructions for Use. Reference the SteriTite Instructions for Use for the recommended use of disposable sterilization accessories. (i.e. Chemical Indicators, Load Cards, Tamper Evident Seals, etc.) with the MediTray Insert Basket and the SteriTite Container. PACKAGING METHODS! 43 English

44 The packaging method recommended for the sterilization, sterile storage and transport of the ME1000 Surgical Impactor is described below and specified in the PACKAGING INSTRUCTIONS section of these Instructions for Use. No other packaging method is recommended for use with this product. Packaging Method: SteriTite Container w/meditray : SteriTite Full Size Sealed Container w/lid and w/etched MED Logo with enclosed MediTray Full Size Insert Basket Preconfigured for MED. See Figures 56,57 For non-sterile storage of the Surgical Impactor, ONLY use the SteriTite Container w/meditray packaging method or if in good condition, the original product packaging. Contact your MED Sales Representative if replacement product packaging is needed. PACKAGING INSTRUCTIONS Recommended Packaging for Surgical Impactor: Case Medical SteriTite Full Size Sealed Container w/lid w/etched MED Logo REF: SC06FG Case Medical SteriTite Disposable Filters Cellulosic 7.5" Dia. 1000/pk. (For use with REF: SC06FG SteriTite Full Size Sealed Container) REF: SCF01 Case Medical MediTray Full Size Insert Basket Preconfigured for MED REF: BSKF06 Packaging Steps for Surgical Impactor: See Figures WARNINGS: ALWAYS make sure the Surgical Impactor has been properly cleaned, lubricated, inspected, functional tested and is completely dry prior to sterile packaging. ONLY load the devices indicated in the TABLE 1 PACKAGING section of these Instructions for Use into the MediTray Insert Basket Preconfigured for MED. NOTE: Reference the SteriTite Instructions for Use for the recommended use of disposable sterilization accessories. (i.e. Chemical Indicators, Load Cards, Tamper Evident Seals, etc.) with the MediTray Insert Basket and the SteriTite Container. 1.Load clean, lubricated and dry Surgical Impactor into its designated place in the MediTray Insert Basket. Prior to loading, place the Anvil in the fully extended position and the Battery Pack Latch in the fully open position. Place the Locking Collar in the Locked Position. Position the Surgical Impactor in the MediTray Insert Basket with the catalog number facing up. Fig. 55! 44 English

45 2.Place the loaded MediTray Insert Basket into the SteriTite Container. SteriTite Sterilization Container Base and Loaded MediTray Insert Basket BSKF06 MediTray Full Size Insert Basket Preconfigured for MED SC06FG SteriTite Full Size Sealed Container Base w/etched MED Logo Fig Place SteriTite Filters into the filter housings on the SteriTite Container Lid as indicated in the Case Medical SteriTite Instructions for Use. 4.Place the SteriTite Container Lid over the SteriTite Container and latch securely. SteriTite Sterilization Container Base, Lid and enclosed MediTray Insert Basket SteriTite Full Size Container Lid for SC06FG w/etched MED Logo BSKF06 MediTray Full Size Insert Basket Preconfigured for MED SC06FG SteriTite Full Size Sealed Container Base w/etched MED Logo Fig. 57 Table 1: Load Plan for MediTray Full Size Insert Basket Preconfigured for MED! 45 English

46 REF Product Description QTY ME1000 SURGICAL IMPACTOR ME1000 STRAIGHT BROACH ADAPTER ME1000 DUAL OFFSET LEFT BROACH ADAPTER ME1000 DUAL OFFSET RIGHT BROACH ADAPTER 1 LOADED BSKF06 MEDITRAY INSERT BASKET PRECONFIGURED FOR MED ME1000 Straight Broach Adapter ME1000 Dual Offset Left Broach Adapter ME1000 Surgical Impactor ME1000 Dual Offset Right Broach Adapter Fig. 58 Table 2: Dimensions for SteriTite Container and MediTray Insert Basket REF Product Description Dimensions SC06FG BSKF06 STERITITE FULL SIZE SEALED CONTAINER W/LID W/ETCHED MED LOGO MEDITRAY FULL SIZE INSERT BASKET PRECONFIGURED FOR MED 23.5 x 11 x 6 (59.7 x 28 x 15.24cm) 21.3" x 9.7" x 6.0" (54.1 x 24.6 x 15.2cm)! 46 English

INSTRUCTIONS FOR USE (IFU)

INSTRUCTIONS FOR USE (IFU) ! INSTRUCTIONS FOR USE (IFU) ME1000 BROACH ADAPTERS Dual Offset Left Broach Adapter 1010-01-105 Straight Broach Adapter 1010-01-101 Dual Offset Right Broach Adapter 1010-01-104 Rx ONLY English (EN) TABLE

More information

User s Manual. Battery Powered Driver. For tapping and inserting screws into varied bone densities.

User s Manual. Battery Powered Driver. For tapping and inserting screws into varied bone densities. User s Manual Battery Powered Driver. For tapping and inserting screws into varied bone densities. Table of Contents Introduction Battery Powered Driver 2 Operating Instructions Assembly 3 Screw Insertion

More information

PM-X IR SURGICAL POWER & ACCESSORIES

PM-X IR SURGICAL POWER & ACCESSORIES PM-X00-770-IR-12.13 SURGICAL POWER & ACCESSORIES Table of Contents Introduction Page 4 Intended Use Page 4 Warning Page 4 Caution Page 5 Notes Page 5 Explanation of Symbols Page 5 Features Page 6 Cleaning

More information

QUANTUM ARM POSITIONER IFU (INSTRUCTIONS FOR USE) The Quantum Arm Positioning System

QUANTUM ARM POSITIONER IFU (INSTRUCTIONS FOR USE) The Quantum Arm Positioning System QUANTUM ARM POSITIONER IFU (INSTRUCTIONS FOR USE) The Quantum Arm Positioning System Quantum OPS Inc. PO Box 4808 Chattanooga, TN 37405 423-805-7999 888-483-9879 contact@quantumops.com v.2019 Copyright

More information

Operator s Manual. Medium-Duty Electric Slicers ENGLISH. Item Model Description Drive Peak HP Voltage Amps Hz Plug

Operator s Manual. Medium-Duty Electric Slicers ENGLISH. Item Model Description Drive Peak HP Voltage Amps Hz Plug Medium-Duty Electric Slicers 40950 40951 Item Model Description Drive Peak HP Voltage Amps Hz Plug 40950 SLM250/S 10 Knife Medium-Duty Slicer 1/3 2.5 Belt 120 60 5-15P 40951 SLM300P/S 12 Knife Medium-Duty

More information

Heyinovo. Disposable Biopsy Valves User Manual. Technical Publications. Document No: WI-RD-13-M07, Rev. B

Heyinovo. Disposable Biopsy Valves User Manual. Technical Publications. Document No: WI-RD-13-M07, Rev. B Heyinovo Disposable Biopsy Valves User Manual Technical Publications Document No: WI-RD-13-M07, Rev. B Copyright By Wilson Instruments (SHA) Co., Ltd. Regulatory Requirement This product complies with

More information

Symmetry Surgica l QUAD-LOCK Sterilization Container System

Symmetry Surgica l QUAD-LOCK Sterilization Container System Symmetry Surgica l QUAD-LOCK Sterilization Container System Symmetry Surgical Inc. 3034 Owen Drive Antioch, TN 37013 USA 1-800-251-3000 Fax: 1-615-964-5566 www.symmetrysurgical.com EC REP Symmetry Surgical

More information

Items that are not wrapped or packaged correctly can allow contamination and therefore present a risk of infection to the patient.

Items that are not wrapped or packaged correctly can allow contamination and therefore present a risk of infection to the patient. Ref. : 1 of 5 PURPOSE AND SCOPE To ensure all staff working in the dental practice responsible for reprocessing of medical instruments and equipment are aware of the correct packaging and wrapping that

More information

LDR Brachytherapy Seed Sterilization and Sorting Tray

LDR Brachytherapy Seed Sterilization and Sorting Tray LDR Brachytherapy Seed Sterilization and Sorting Tray Table of Contents General Precautions... 2 Customer Responsibility... 3 Description... 4 Features and Specifications... 5 Operation... 6 Changing Covers

More information

Bard. BI-DIRECTIONAL and KELLY-WICK TUNNELERS. Instructions For Use. PK Rev. 1 04/06

Bard. BI-DIRECTIONAL and KELLY-WICK TUNNELERS. Instructions For Use. PK Rev. 1 04/06 Bard BI-DIRECTIONAL and KELLY-WICK TUNNELERS Instructions For Use PK0114100 Rev. 1 04/06 Bard Bi-Directional and Kelly-Wick Tunnelers Caution: Federal (USA) law restricts this device to sale by or on the

More information

THE EDGE12 SLICER. MODEL EDGE12-2 (220/50/1 voltage) FORM (July 2012) EDGE12 SLICER

THE EDGE12 SLICER. MODEL EDGE12-2 (220/50/1 voltage) FORM (July 2012) EDGE12 SLICER EDGE12 SLICER THE EDGE12 SLICER MODEL EDGE12-1 (120/60/1 voltage) EDGE12-2 (220/50/1 voltage) ML-136246 ML-136247 701 S. RIDGE AVENUE TROY, OHIO 45374-0001 937 332-3000 www.hobartcorp.com FORM 35283 (July

More information

Retractor Kit Rev: D. Instruction For Use NEXT ORTHOSURGICAL 3270 CORPORATE VIEW, SUITE A VISTA CA /1/14

Retractor Kit Rev: D. Instruction For Use NEXT ORTHOSURGICAL 3270 CORPORATE VIEW, SUITE A VISTA CA /1/14 Instruction For Use NEXT ORTHOSURGICAL 3270 CORPORATE VIEW, SUITE A VISTA CA 92081 8/1/14 Summary 1. User manual... 3 1.1. Presentation... 3 1.1.1. Destination... 3 1.1.2. Surgical context... 3 1.1.3.

More information

Meat Slicer INSTRUCTION MANUAL CAUTION! ONE YEAR LIMITED WARRANTY

Meat Slicer INSTRUCTION MANUAL CAUTION! ONE YEAR LIMITED WARRANTY ONE YEAR LIMITED WARRANTY INSTRUCTION MANUAL The original registered owner of this product should contact SKYFOOD EQUIPMENT LLC at 305-868-16 for any warranty problems or service. SKYFOOD EQUIPMENT LLC

More information

Motorized retractable awning

Motorized retractable awning Motorized retractable awning Model 95295 Set up And Operating Instructions Diagrams within this manual may not be drawn proportionally. Due to continuing improvements, actual product may differ slightly

More information

Instructions for Use A-dec Quick-Disconnect Syringes

Instructions for Use A-dec Quick-Disconnect Syringes Instructions for Use A-dec Quick-Disconnect Syringes Contents Basic Operation...1 Adjustments... 4 Asepsis... 6 Specifications, Warranty, and Product Service... 8 Basic Operation Water, Air, and Spray

More information

SAFETY AND OPERATING MANUAL

SAFETY AND OPERATING MANUAL SAFETY AND OPERATING MANUAL General Safety Rules WARNING! Read all instructions. Failure to follow all instructions listed below may result in electric shock, fire and/ or serious injury. The term power

More information

electric knife sharpener

electric knife sharpener electric knife sharpener USER GUIDE Now that you have purchased a Kitchen Living product you can rest assured in the knowledge that as well as your year parts and labour guarantee you have the added peace

More information

Installation and User s Manual 12 x 10 MOTORIZED AWNING

Installation and User s Manual 12 x 10 MOTORIZED AWNING 12 x 10 MOTORIZED AWNING Installation and User s Manual 12 x 10 MOTORIZED AWNING 088-1763-0 Stop Please read and understand this manual before any assembly or use of this product. Before beginning assembly

More information

Pediatric Lower Body Warming Blanket Model B203 Instructions for Use

Pediatric Lower Body Warming Blanket Model B203 Instructions for Use Pediatric Lower Body Warming Blanket Model B203 Instructions for Use Manufactured by: Augustine Temperature Management, LLC 6581 City West Parkway Eden Prairie, MN 55344 USA TEL 952.465.3500 FAX 952.465.3501

More information

WARMING BLANKETS INSTRUCTIONS FOR USE

WARMING BLANKETS INSTRUCTIONS FOR USE WARMING BLANKETS INSTRUCTIONS FOR USE INTRODUCTION HotDog Warming Blankets and optional Disposable Sheets are part of the HotDog Patient Warming System. These instructions apply to the following catalog

More information

MAX Series 9514 OPERATOR S MANUAL

MAX Series 9514 OPERATOR S MANUAL MAX Series 9514 OPERATOR S MANUAL Persons under age 18 are not permitted to operate or have accessibility to operate this equipment per U.S. Dept. Of Labor Employment Standards Administration Fact Sheet

More information

User Guide. Tripod. flowtech 75 Tripod. Part No. V

User Guide. Tripod. flowtech 75 Tripod.  Part No. V User Guide Tripod flowtech 75 Tripod Part No. V4150-0001 EN www.vinten.com TM Copyright 2017 All rights reserved. Original Instructions: English All rights reserved throughout the world. No part of this

More information

USER INSTRUCTIONS. Applicable to Lanyard : REF. NO. : ,

USER INSTRUCTIONS. Applicable to Lanyard : REF. NO. : , USER INSTRUCTIONS Applicable to Lanyard : REF. NO. : 10185600, 10185613 Manufacturer : MSA Europe GmbH, Schlüsselstr. 12, CH- 8645 Rapperswil-Jona Switzerland Body : SATRA Technology Centre, Wyndham Way,

More information

STERILE CONTAINER SYSTEMS

STERILE CONTAINER SYSTEMS »» Extra Long Size EXTRA LONG SIZE 280 mm 710 mm h Specificiations Lid (Perforated) Lid (Perforated) Sizes Box (Non Perforated) Box (Perforated) TK6-2010 TK6-2040 TK6-2011 TK6-2041 h=120 mm TK6-2012 TK6-2013

More information

1000M MANUAL Signature Series Slicer OPERATORS MANUAL

1000M MANUAL Signature Series Slicer OPERATORS MANUAL 1000M MANUAL Signature Series Slicer OPERATORS MANUAL Persons under age 18 are not permitted to operate or have accessibility to operate this equipment per U.S. Dept. Of Labor Employment Standards Administration

More information

OWNER S MANUAL. Item #14103

OWNER S MANUAL. Item #14103 TRI-BALL HITCH WITH CLEVIS OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in serious

More information

300 ft. 5/8 Hose wagon

300 ft. 5/8 Hose wagon 300 ft. 5/8 Hose wagon Model 95956 Assembly And Operation Instructions Due to continuing improvements, actual product may differ slightly from the product described herein. (Garden hose is not included).

More information

Manual Awning. Assembly Instructions. Product No Toll-free:

Manual Awning. Assembly Instructions. Product No Toll-free: Manual Awning Product No. 088-30- Instructions Toll-free: -877-483-679 IMPORTANT: Please read this manual carefully before beginning assembly of this product. Keep this manual for future reference. 3 Table

More information

STERILE CONTAINER SYSTEMS

STERILE CONTAINER SYSTEMS »» Extra Long Size EXTRA LONG SIZE 280 mm 710 mm h h 1 Specificiations Lid (with barrier) Lid (with barrier) Box (Non Perforated) Box (with barrier) TK7-2010 TK7-2040 TK7-2011 TK7-2041 TK7-2012 TK7-2013

More information

S E L E C T I O N. Abdominal Crunch. User manual

S E L E C T I O N. Abdominal Crunch. User manual and S E L E C T I O N T H E S T R E N G T H E V O L U T I O N User manual and and The identification plate of the and manufacturer, affixed along side the seat on the frame of the weight stack, gives the

More information

FORM (Apr. 2006)

FORM (Apr. 2006) 2212 Slcier MODEL 2212 SLICER MODEL 2212 ML-136132 701 S. RIDGE AVENUE TROY, OHIO 45374-0001 937 332-3000 www.hobartcorp.com FORM 35215 (Apr. 2006) Installation, Operation, and Care of MODEL 2212 Slicer

More information

Instructions for Use: Duodenoscope Sampling Kit

Instructions for Use: Duodenoscope Sampling Kit Instructions for Use: Duodenoscope Sampling Kit Brand Name of Product Generic Name of Product Product Code Number(s) Intended Use Duodenoscope Sampling Kit Endoscope sampling and culturing kit CK-250,

More information

INSTALLATION & MAINTENANCE MANUAL SERIES 9000 MODELS 9150/9300 SERIES 9000 VALVES WITH V BALL

INSTALLATION & MAINTENANCE MANUAL SERIES 9000 MODELS 9150/9300 SERIES 9000 VALVES WITH V BALL BRIEF INTRODUCTION Split body (2-piece) valve allows easy replacement of gasket, seal, and seats without special tools. Series 90D ball valves use floating ball design. Induced by the line pressure the

More information

User Guide. Tripod. flowtech 75 Tripod. Part No. V

User Guide. Tripod. flowtech 75 Tripod.  Part No. V User Guide Tripod flowtech 75 Tripod Part No. V4150-0001 EN www.vinten.com TM Copyright 2017 All rights reserved. Original Instructions: English All rights reserved throughout the world. No part of this

More information

SERVICE MANUAL- M126

SERVICE MANUAL- M126 SERVICE MANUAL- MANUAL CAN OPENER- MODEL # 2 1 Model #2 Can Opener Assembly Procedure The Model #2 can opener will be assembled according to the following procedure. I. Model #2 Handle and Arbor Assembly

More information

SURGICAL TECHNIQUE. Legend Acetabular Shell

SURGICAL TECHNIQUE. Legend Acetabular Shell SURGICAL TECHNIQUE Legend Acetabular Shell The following technique is a general guide for instrumentation of the Legend Acetabular Shell. It is expected that the surgeon is already familiar with the fundamentals

More information

ADC Satin TM Standard, Fiber Optic and Disposable Laryngoscope Blades and Handles

ADC Satin TM Standard, Fiber Optic and Disposable Laryngoscope Blades and Handles ADC Satin TM Standard, Fiber Optic and Disposable Laryngoscope Blades and Handles Directions for Use Thank you for choosing an ADC Satin Laryngoscope. Device Description and Intended Use A Laryngoscope

More information

Operation Manual SUPER BEAST Multi-tool SC-11M-10

Operation Manual SUPER BEAST Multi-tool SC-11M-10 Operation Manual SUPER BEAST Multi-tool SC-11M-10 Technical Specifications General Safety/Operating Instructions Attaching/Changing the Head Using the SUPER BEAST as a Cutter Using the SUPER BEAST as a

More information

FagronLab FG. Semi-automatic capsule filling machine. User Manual

FagronLab FG. Semi-automatic capsule filling machine. User Manual Semi-automatic capsule filling machine User Manual Table of contents 1. Introduction... 4 2. Service.. 4 3. Warranty... 4 4. Overview... 5 System features... 5 FagronLab FG-1 Filler Base & Orienter Base...

More information

JARVIS. Model 70 Airsnip Air Powered Scissors

JARVIS. Model 70 Airsnip Air Powered Scissors Air Powered Scissors EQUIPMENT SELECTION... Ordering No. Airsnip Model 70... 4019009 Air Filter/Regulator/Lubricator.. 3022003 Balancer... 1350084 Power--Pak (250 PSI)... 4026016 TABLE OF CONTENTS... Page

More information

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS GENUINE PARTS INSTALLATION INSTRUCTIONS 1. DESCRIPTION: Hatch Tent 2. APPLICATION: See dealer application chart 3. PART NUMBER: 999T7 XR100 and 999T7 XY100 4. KIT CONTENTS: Item Qty Description A 1 Tent

More information

MODEL # 1 MANUAL CAN OPENER

MODEL # 1 MANUAL CAN OPENER REV.B SERVICE MANUAL MODEL # 1 MANUAL CAN OPENER Model #1 Can Opener Assembly Procedure The Model #1 can opener will be assembled according to the following procedure. I. Model #1 Handle and Arbor Assembly

More information

Retractable Lance Instructions

Retractable Lance Instructions Retractable Lance Instructions BETE Fog Nozzle Inc. www.bete.com Safety The following symbols are used to call the reader s particular attention to areas of concern: General Warning: This symbol alerts

More information

For the diagnosis of Hirschsprung s Disease

For the diagnosis of Hirschsprung s Disease For the diagnosis of Hirschsprung s Disease Suction Rectal Biopsy System Suitable for use with: Neonates - Adolescents - Adults World s #1 System Patent No US 7,479,117 B2 Controlled Consistent Rectal

More information

NORTHFIELD CORPORATION 1870 COMMERCE DR. DE PERE, WI UNITED STATES

NORTHFIELD CORPORATION 1870 COMMERCE DR. DE PERE, WI UNITED STATES 1 MANUFACTURER: MODEL: NORTHFIELD CORPORATION 1870 COMMERCE DR. DE PERE, WI 54115 UNITED STATES NORTHFIELD CLS SPECIFICATIONS: VOLTAGE: FREQUENCY: CURRENT: AIR: 115VAC-230VAC 50/60Hz 5 AMP 80 PSI (for

More information

Instruction Manual please read before use

Instruction Manual please read before use Instruction Manual please read before use XL Burner (patent pending) 307 Tchoupitoulas Street, Suite 100 New Orleans, LA 70130 p : 888.596.6960 504.596.6960 e : office@brasafire.com www.brasa.co Item Identification

More information

Operator s Manual. Wedgemaster 606N 808N 808SG ENGLISH

Operator s Manual. Wedgemaster 606N 808N 808SG ENGLISH ENGLISH Wedgemaster 606N 808N 808SG Thank you for purchasing this Vollrath Food Processing Equipment. Before operating the equipment, read and familiarize yourself with the following operating and safety

More information

ALPINE SLIDEOUT COVER

ALPINE SLIDEOUT COVER INSTALLATION MANUAL ALPINE SLIDEOUT COVER RV Read this manual before installing or using this product. Failure to follow the instructions and safety precautions in this manual can result in personal injury

More information

Product instruction manual Ream Cutting Systems RE3943, RE3946, RE3947, RE3971, RE3952E

Product instruction manual Ream Cutting Systems RE3943, RE3946, RE3947, RE3971, RE3952E Product instruction manual Ream Cutting Systems RE3943, RE3946, RE3947, RE3971, RE3952E The Trimfast Ream Cutters are reliable, high performance cutters that will give you the results you need quickly

More information

TUCSON PATIO AWNING USE & MAINTENANCE

TUCSON PATIO AWNING USE & MAINTENANCE TUCSON PATIO AWNING USE & MAINTENANCE Toll Free 877-792-1775 Fax 877-792-0031 swsuncontrolpro.com sales@swsuncontrol.com 1 INTRODUCTION...3 2 SAFETY...3 2.1 General safety information... 3 2.2 Safety devices...

More information

Tuke Saw System. Operating Manual. Safe Navigation Through Soft Tissue

Tuke Saw System. Operating Manual. Safe Navigation Through Soft Tissue Tuke Saw System Operating Manual Safe Navigation Through Soft Tissue Contents Contents Section 1 Introduction 3 Section 2 Special features 4 Section 3 Operation 5 Section 4 Use of blades and brushes 9

More information

ADDED PROTECTION FOR SLIDE-OUT ROOMS. Installation. Introduction... 2 Product Overview...2 Component Checklist...3

ADDED PROTECTION FOR SLIDE-OUT ROOMS. Installation. Introduction... 2 Product Overview...2 Component Checklist...3 CAREFREE SIDEOUT KOVER II RV Installation ADDED PROTECTION FOR SLIDE-OUT ROOMS TABLE OF CONTENTS SOKII Shown with Full Case Option Introduction... 2 Product Overview...2 Component Checklist...3 Installation...

More information

1 Pan - 6 Ways to Cook! Works in the Oven & On All Stovetops! electric gas ceramic induction. Broil Bake Sauté Fry Steam Braise

1 Pan - 6 Ways to Cook! Works in the Oven & On All Stovetops! electric gas ceramic induction. Broil Bake Sauté Fry Steam Braise *only use non-metal utensils 1 Pan - 6 Ways to Cook! Broil Bake Sauté Fry Steam Braise Non-stick Cerami-Tech Ceramic Coating Even Heating Good for All Surfaces Easy Cleaning Durable, Stain Resistant Coating

More information

Application Tooling Specification

Application Tooling Specification Order No. 200218-3100 FEATURES Application Tooling Specification A full cycle ratcheting hand tool ensures complete crimps Long handles for comfortable crimping with reduced crimping force and a safety

More information

PS-10 PS-12D PS-12. Primo Slicer Manual FOR OPERATOR-DO NOT DISCARD

PS-10 PS-12D PS-12. Primo Slicer Manual FOR OPERATOR-DO NOT DISCARD PS-10 PS-12D PS-12 Primo Slicer Manual FOR OPERATOR-DO NOT DISCARD TABLE OF CONTENTS 1. BRIEF INTRODUCTION 2. OPERATION 3. SLICING 4. MAINTENANCE AND CLEANING 5. TROUBLESHOOTING IMPORTANT The operator

More information

Blue Series Hot Melt Hose with RediFlext II Hanger System

Blue Series Hot Melt Hose with RediFlext II Hanger System Instruction Sheet P/N 1124594_01 Blue Series Hot Melt Hose with RediFlext II Hanger System Safety WARNING! Allow only personnel with appropriate training and experience to operate or service the equipment.

More information

BR B Bike Rack Instructions

BR B Bike Rack Instructions IMPORTANT NOTE: It is common that vehicles with lids and hatches have a single rear wiper. It is important that you make yourself aware that the wiper cannot be used while the rack is attached to the vehicle.

More information

CRD120SC TUBING CUTTER

CRD120SC TUBING CUTTER CRD120SC TUBING CUTTER OPERATIONS MANUAL 1 VERSION 2.1 LAST EDITED 05.01.2018 cleanroomdevices.com Table of Contents Table of Contents....2 1.0 General Product & Safety Information... 3 1.1 Product Information

More information

INSTRUCTION SHEET TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM PN S: , ,

INSTRUCTION SHEET TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM PN S: , , INSTRUCTION SHEET TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM PN S: 22-797010, 22-797020, 22-796161 COMPONENTS: 1. TVS-100, PN 22-797010: 47 60 (119 cm 152 cm) stroke TwistLock Vehicle Stabilizer. Component

More information

4A1-114/114KR & 4A1-2 MANUAL TENSIONERS

4A1-114/114KR & 4A1-2 MANUAL TENSIONERS M L. C O O N AL TO 4A1-114/114KR & 4A1-2 W W W.T R AD IT IO MANUAL TENSIONERS READ THESE INSTRUCTIONS CAREFULLY. FAILURE TO FOLLOW THESE INSTRUCTIONS CAN RESULT IN SEVERE PERSONAL INJURY. GENERAL SAFETY

More information

User Manual. 110 Cup (55 Cup Raw) Rice Cooker. Model: 177GRCLP, 177GRCNAT 10/2017. Please read and keep these instructions. Indoor use only.

User Manual. 110 Cup (55 Cup Raw) Rice Cooker. Model: 177GRCLP, 177GRCNAT 10/2017. Please read and keep these instructions. Indoor use only. 110 Cup (55 Cup Raw) Rice Cooker Model: 177GRCLP, 177GRCNAT 10/2017 Please read and keep these instructions. Indoor use only. www.avantcoequipment.com 1 NOTE: Save these instructions for future reference.

More information

Eaton ET Hydraulic Hose Saw INSTRUCTION MANUAL

Eaton ET Hydraulic Hose Saw INSTRUCTION MANUAL Eaton ET9100-07-110 Hydraulic Saw INSTRUCTION MANUAL INSTRUCTION MANUAL Eaton ET9100-07-110 Hydraulic Saw Table of Contents Introduction... 3 Old Method... 3 New Eaton Method... 3 Operation... 4 Operation

More information

SB3T. Roll-InBuddy + Tilt. Dignity + Ease + Comfort. Product features include:

SB3T. Roll-InBuddy + Tilt. Dignity + Ease + Comfort. Product features include: SB3T Roll-InBuddy + Tilt The revolutionary tilt option incorporated into the Roll-InBuddy Tilt allows for pressure release as well as a more thorough bathing experience for the user and easier reach for

More information

Capri. Instruction Manual. Balcony Umbrella x 6.93 Rectangle / SSBU-5X7RT5T

Capri. Instruction Manual. Balcony Umbrella x 6.93 Rectangle / SSBU-5X7RT5T Capri Balcony Umbrella 4.95 x 6.93 Rectangle / SSBU-5X7RT5T Instruction Manual Thank you for purchasing SimplyShade s Capri Balcony Umbrella. Please remove all contents from the package, inspect and review

More information

OWNER'S MANUAL FREEDOM AWNINGS

OWNER'S MANUAL FREEDOM AWNINGS OWNER'S MANUAL FREEDOM AWNINGS RV PATIO, WINDOW AND OTD AWNINGS Freedom Patio Awning Freedom Window and OtD Awnings Special Note: This manual provides a basic description of the product and the standard

More information

RUPHIS Blade Clamp. Instruction Manual

RUPHIS Blade Clamp. Instruction Manual RUPHIS Blade Clamp P/N 1006800 NSN 1615-01-562-8156 Instruction Manual AH-64D UH-60A/L UH-60M S-70i S-92 REVISION AUTHORIZATION / RECORD These instructions are primarily for inspection of the RUPHIS Blade

More information

INSTRUCTIONS INSTALLATION UNIVERSAL SERIES PATIO AWNING HARDWARE , , ,

INSTRUCTIONS INSTALLATION UNIVERSAL SERIES PATIO AWNING HARDWARE , , , RECORD THIS INFORMATION FOR FUTURE REFERENCE: Model Number Serial Number Hardware Model Number Hardware Serial Number Date Purchased Retailer / Qualified Installer INSTALLATION INSTRUCTIONS UNIVERSAL SERIES

More information

Levels & Measuring Tools

Levels & Measuring Tools Levels & Measuring Tools Quick, easy and accurate measurements can mean the difference between a job well done and one that misses the mark. Klein offers a variety of levels and other measuring devices

More information

NEWMAR SERVICE SCHOOL

NEWMAR SERVICE SCHOOL NEWMAR SERVICE SCHOOL TRAINING INFORMATION GUIDELINE FOR FEBRUARY 2013 OUR PRODUCTS: NOVA DUAL PITCH AWNING G-2000/ G-1500 2 P a g e G-2085 G-5000 3 P a g e G-LINKS 4 P a g e NOVA/ G-2000/ G-1500 BASIC

More information

Cordless Electric Teakettle. Model 685

Cordless Electric Teakettle. Model 685 Cordless Electric Teakettle Model 685 Important Safeguards When using electrical appliances, basic safety precautions should always be followed, including the following: 1. Read all instructions. 2. Do

More information

INSTALLATION GUIDE INTRODUCTION

INSTALLATION GUIDE INTRODUCTION INSTALLATION GUIDE for for HotEdge Rail Rail INTRODUCTION HotEdge Rail is a patent pending one-piece roof edge ice melt systems engineered to easily install on standing seam, pro panel, and corrugated

More information

Wireless Wind Sensor Installation and Operation Instructions

Wireless Wind Sensor Installation and Operation Instructions WARNINGS: RETRACTABLE AWNINGS For Technical Support visit us at www.sunsetter.com/ownerscorner or Call Toll Free 800-670-7071 Fax 877-224-4944 Wireless Wind Sensor Installation and Operation Instructions

More information

Kontrol Kube Advanced Owners Manual

Kontrol Kube Advanced Owners Manual Mobile Containment. Simplified. TM Kontrol Kube Advanced Owners Manual KONTROL KUBE www.kontrolkube.com 800.343755 1 IMPORTANT SAFETY INFORMATION SAVE THESE INSTRUCTIONS. CAREFULLY READ AND FOLLOW THESE

More information

I N ST R UC T I ON MODELS 1612E, 1712E & 1712RE SLICERS. FORM Rev. B (10-97) 1712RE 701 S. RIDGE AVENUE TROY, OHIO 45374

I N ST R UC T I ON MODELS 1612E, 1712E & 1712RE SLICERS. FORM Rev. B (10-97) 1712RE 701 S. RIDGE AVENUE TROY, OHIO 45374 I N ST R UC 1712E T I ON MODELS 1612E, 1712E & 1712RE SLICERS S ML-104546 ML-104547 ML-104548 1612E 1712E 1712RE 701 S. RIDGE AVENUE TROY, OHIO 45374 FORM 19238 Rev. B (10-97) TABLE OF CONTENTS GENERAL......................................................................

More information

OWNER'S MANUAL CAMPOUT TABLE OF CONTENTS

OWNER'S MANUAL CAMPOUT TABLE OF CONTENTS OWNER'S MANUAL CAMPOUT RV Before operating the awning, carefully review the Owner's Manual. The manual contains important safety information, detailed operating instructions, common maintenance procedures

More information

Laboratory Safety. A practical approach.

Laboratory Safety. A practical approach. Laboratory Safety A practical approach. 1 SDS Management All SDS Material is filed in a central location An additional copy is placed in the room where the item is stored Central location files are permanent

More information

Instruction Manual please read before use

Instruction Manual please read before use Instruction Manual please read before use Slim Burner Insert (patent pending) 307 Tchoupitoulas Street, Suite 100 New Orleans, LA 70130 p : 888.596.6960 504.596.6960 e : office@brasafire.com www.brasafire.com

More information

Capacity kw. Thermal turndown from specified maximum capacity 8:1 8:1

Capacity kw. Thermal turndown from specified maximum capacity 8:1 8:1 3-18.3-5 Specifications of OXY-THERM TITAN burners Typical burner data Fuel: natural gas at 15 C with 10.9 kwh/nm 3 HHV - sg = 0.6 Oxygen: 90-100% purity Stated pressures are indicative. Actual pressures

More information

TN5000 Titan Manual Rotary Microtome

TN5000 Titan Manual Rotary Microtome User Manual TN5000 Titan Manual Rotary Microtome Read Prior to Operation! Always keep this manual near the unit! WP47617 06.02.15 User Manual TN5500: Titan 888.708.5233 Info@TannerScientific.com www.tannerscientific.com

More information

SINGLE BURNER STOVE IMPORTANT OPERATING INSTRUCTIONS COOKING & STORAGE POSITION PARTS IDENTIFICATION

SINGLE BURNER STOVE IMPORTANT OPERATING INSTRUCTIONS COOKING & STORAGE POSITION PARTS IDENTIFICATION SINGLE BURNER STOVE OPERATING INSTRUCTIONS PARTS IDENTIFICATION Electronic Lighter Ignites the burner. Liquid Guard Holds liquid overflow. Windshield Prevents flame from being extinguished by wind. Ensures

More information

TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM INSTRUCTIONS

TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM INSTRUCTIONS TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM 22-797010, 22-797020, 22-796161 INSTRUCTIONS COMPONENTS: 1. TVS-100, PN 22-797010: 47 60 (119 cm 152 cm) stroke TwistLock Vehicle Stabilizer. Component weight,

More information

AUTO REWIND HOSE REEL USER MANUAL

AUTO REWIND HOSE REEL USER MANUAL AUTO REWIND HOSE REEL USER MANUAL Models: US3025 IMPORTANT: READ BEFORE USE AND KEEP FOR FUTURE REFERENCE Welcome to trouble-free watering at its best! The Hoselink Auto Rewind Hose Reel is manufactured

More information

Reprocessing Instructions and Inspecton Criteria

Reprocessing Instructions and Inspecton Criteria Reprocessing Instructions and Inspecton Criteria DEVICE DESCRIPTION: The SteriTite universal container is a rigid, reusable, sealed medical sterilization packaging system validated for steam and low temperature

More information

Table of Contents. Safety... 2 Specifications... 4 Assembly Maintenance... 6 Parts List... 7 Warranty... 8

Table of Contents. Safety... 2 Specifications... 4 Assembly Maintenance... 6 Parts List... 7 Warranty... 8 Table of Contents Safety... Specifications... 4 Assembly... 4 Maintenance... 6 Parts List... 7 Warranty... 8 WARNING SYMBOLS AND DEFINITIONS This is the safety alert symbol. It is used to alert you to

More information

Lit Table Top Firepit Bio-Ethanol Fireplace. User Manual. Model: GF301650

Lit Table Top Firepit Bio-Ethanol Fireplace. User Manual. Model: GF301650 Lit Table Top Firepit Bio-Ethanol Fireplace User Manual Model: GF301650 1 Moda Flame Lit Table Top Firepit Bio-Ethanol Fireplace Included in delivery (Parts List): A. Fireplace body 1pc B. Glass 1pc C.

More information

Installation and Maintenance Manual

Installation and Maintenance Manual Freestanding Gas Stove MODEL: PGS2005 GPEBB20R GPEBW20R Installation and Maintenance Manual Warning Maintenance products should be carried out by professional and technical personnel with relevant qualification,

More information

Sterile Processing Department Solutions CLEANING POWER IN YOUR HANDS INSTRUMENT CLEANING BRUSHES. One Integrated Approach to Healthcare.

Sterile Processing Department Solutions CLEANING POWER IN YOUR HANDS INSTRUMENT CLEANING BRUSHES. One Integrated Approach to Healthcare. Sterile Processing Department Solutions CLEANING POWER IN YOUR HANDS INSTRUMENT CLEANING BRUSHES One Integrated Approach to Healthcare. Hold the cleaning power in your hands Harness the cleaning power

More information

OUTDOOR SERIES Mosquito protection you don t spray on Repels mosquitoes, black flies and flying insects Provides 15 x15 ft of repellency, the size of

OUTDOOR SERIES Mosquito protection you don t spray on Repels mosquitoes, black flies and flying insects Provides 15 x15 ft of repellency, the size of OUTDOOR SERIES Mosquito protection you don t spray on Repels mosquitoes, black flies and flying insects Provides 15 x15 ft of repellency, the size of an average deck LEDs provide safe, ambient light Each

More information

DOWN MANUAL. Aeros Ltd. St. Post-Volinskaya, 5 Kiev, UKRAINE

DOWN MANUAL. Aeros Ltd. St. Post-Volinskaya, 5 Kiev, UKRAINE DOWN HG & PG rescue parachute system MANUAL Aeros Ltd. St. Post-Volinskaya, 5 Kiev, 03061 UKRAINE Tel. +(380 44) 455 41 20 Fax. +(380 44) 455 41 16 E-mail: aerosint@aerosint.kiev.ua http://www.aeros.com.ua

More information

Termination Procedure LC 1.6mm & 2.0mm Connectors: 1060XX Series

Termination Procedure LC 1.6mm & 2.0mm Connectors: 1060XX Series Connector Piece Parts LC Simplex Connector Boot LC Duplex Connector LC Connector CriMp Tulolng LC Duplex Clip Boot 86001-0110, Rev. B USN30MFO/5/2010 Procedure Chart Procedure Tool Required Tool Part Number

More information

Cast Iron Cookware Ustensiles de cuisson en fonte Utensilios de cocina de hierro fundido INSTRUCTIONS INSTRUCTIONS INSTRUCCIONES

Cast Iron Cookware Ustensiles de cuisson en fonte Utensilios de cocina de hierro fundido INSTRUCTIONS INSTRUCTIONS INSTRUCCIONES Cast Iron Cookware Ustensiles de cuisson en fonte Utensilios de cocina de hierro fundido INSTRUCTIONS INSTRUCTIONS INSTRUCCIONES WELCOME TO THE WORLD OF KITCHENAID. We re committed to helping you create

More information

SAVE THESE INSTRUCTIONS.

SAVE THESE INSTRUCTIONS. IMPORTANT SAFEGUARDS When using electrical appliances, basic safety precautions should always be followed including the following: 1. Read all instructions. Every user should read this manual. 2. To protect

More information

CIRRUS AIRPLANE MAINTENANCE MANUAL MODELS SR22 AND SR22T CHAPTER 56-20: CABIN GENERAL. Cabin 56-20: CABIN. 1. General

CIRRUS AIRPLANE MAINTENANCE MANUAL MODELS SR22 AND SR22T CHAPTER 56-20: CABIN GENERAL. Cabin 56-20: CABIN. 1. General CIRRUS AIRPLANE MAINTENANCE MANUAL Cabin CHAPTER 56-20: CABIN GENERAL 56-20: CABIN 1. General The aft passenger windows are manufactured of acrylic and adhesive bonded to the fuselage. Replacement is accomplished

More information

CIRRUS AIRPLANE MAINTENANCE MANUAL

CIRRUS AIRPLANE MAINTENANCE MANUAL CABIN 1. GENERAL The aft passenger windows are manufactured of acrylic and adhesive bonded to the fuselage. Replacement is accomplished by removing the interior trim around the window, removing the sealant

More information

1612P MODEL 1612P SLICER MODEL EXECUTIVE OFFICES 701 RIDGE AVENUE TROY, OHIO FORM (4-95)

1612P MODEL 1612P SLICER MODEL EXECUTIVE OFFICES 701 RIDGE AVENUE TROY, OHIO FORM (4-95) 1612P MODEL 1612P SLICER MODEL 1612P ML-104587 EXECUTIVE OFFICES 701 RIDGE AVENUE TROY, OHIO 45374-0001 FORM 19370 (4-95) Installation, Operation, and Care of MODEL 1612P SLICER SAVE THESE INSTRUCTIONS

More information

Phantom Retractor System Overview

Phantom Retractor System Overview Spine Phantom Retractor System Overview Minimally invasive procedures Dual-directional expandable retractor 30 degree pivoting blades Fiber-optic light source for better visualization Key Design Features

More information

PRODUCT MANUAL - M096

PRODUCT MANUAL - M096 PRODUCT MANUAL - M096 MODEL 270 ELECTRIC CAN OPENER 1 I. LABELS L087--CAUTION 2 II. SPECIFICATIONS MODEL NO. 270 POWER 115 VOLT, 1.5 AMP, 50-60HZ REQUIREMENTS 230 VOLT, 0.8 AMP, 50-60 HZ NORMAL SPEED 200-250

More information

INSTRUCTION MANUAL HYDM140

INSTRUCTION MANUAL HYDM140 INSTRUCTION MANUAL HYDM140 INSTRUCTION MANUAL Clean-Cuts HYDM140 Hydraulic Hose Saw Table of Contents Introduction... 3 Old Method... 3 New Clean-Cuts Method... 3 Operation... 4 Operation Procedure...

More information

Operating Instructions

Operating Instructions Operating Instructions High Throughput Dialysis MODEL HTD96b www.htdialysis.com HTD96b : Operating Instructions Components Universal Base Unit Teflon block assembly: 9 Teflon bars, labeled sequentially

More information

EB300. Ethanol Burner. User s Manual Installation Instructions

EB300. Ethanol Burner. User s Manual Installation Instructions EB300 Ethanol Burner User s Manual Installation Instructions EB300 Burner Burner Lid User s Manual What s in the box: 1pc 1pc 1pc Installation Instructions EB300 - User s Manual The Ignis Ethanol Burner

More information