The ULM CASE III Emergency Case. Description and instructions for use

Similar documents
S E L E C T I O N. Abdominal Crunch. User manual

Product instruction manual Ream Cutting Systems RE3943, RE3946, RE3947, RE3971, RE3952E

Leica CE/ Leica CN. Knife holder

Minibar User Manual MBA 4000 W

2522 and 2523 Fiber Optic Splice Organizer Tray

INSTANT GARAGE MODEL NO: CIG81224 ASSEMBLY INSTRUCTIONS PART NO: ORIGINAL INSTRUCTIONS

t420 Barbecue Assembly Manual

JOINTED ARM AWNING SWINGLINE CASSETTE

Safety. Set up and align scale. Congratulations

Installation and User s Manual 12 x 10 MOTORIZED AWNING

Operator s Manual. Medium-Duty Electric Slicers ENGLISH. Item Model Description Drive Peak HP Voltage Amps Hz Plug

Rhino-Rack SUNSEEKER Awning Black

AFC-50 Automatic French Fry Cutter Instruction Manual

33100 LH - Batwing Awning RH - Batwing Awning

AEROtube ET-Holder Laryngoscopes Laryngeal masks Endotracheal tubes Oropharyngeal tubes

CRD120SC TUBING CUTTER

SINGLE BURNER STOVE IMPORTANT OPERATING INSTRUCTIONS COOKING & STORAGE POSITION PARTS IDENTIFICATION

HANG GLIDER (RECHARGEABLE)

Kontrol Kube Advanced Owners Manual

RAFTER VI. Installation and Operation CAREFREE WITH AUTOMATIC AWNING SUPPORT. RV Accessory PRODUCT OVERVIEW

jointed arm awning _EN

Installation & Operation Manual VHM Variable Height Series Arms with Fixed Angle Front End

NM-11 INSTRUCTION MANUAL WARNING!! TO REDUCE THE RISK OF INJURY, USER MUST READ AND UNDERSTAND THIS INSTRUCTION MANUAL. ORIGINAL INSTRUCTIONS

Urea/Adblue Hose Reel

TN5000 Titan Manual Rotary Microtome

300 ft. 5/8 Hose wagon

Read all instructions and warnings before using this fireplace. WARNING

280 Propane Barbecue Assembly Manual

MYRIAD Banner Stand is a trademark of Skyline Exhibits. Patent Pending PN32294-B. MYRIAD Banner Stand

USER MANUAL BMU WORKCAGE. CONFORM TO THE MACHINE DIRECTIVE 2006/42/EC and to EN1808 (1999)

MMSlog MULTIFUNCTIONAL TROLLEY

Important Information

MALTE. the compact and versatile posterior-walker. Instructions for use

BR B Bike Rack Instructions

Insruction manual Dukki

BERYL Service Guide. Travel Adjust OTT Change

FEATURES & BENEFITS OF THE TOW PRO LITE :

Capri. Instruction Manual. Balcony Umbrella x 6.93 Rectangle / SSBU-5X7RT5T

2 Burner ASSEMBLY MANUAL

GLM300 Semi-auto Slicer

HARD DOOR UPGRADE KIT

Installation Instructions

ROOF-TOP SAFARI TENT. ASSEMBLY & CARE INSTRUCTIONS Part No ENSURE YOU READ AND UNDERSTAND THESE INSTRUCTIONS PRIOR TO INSTALLATION

Maintenance Instructions and Directions for Use Wind Support

1612P MODEL 1612P SLICER MODEL EXECUTIVE OFFICES 701 RIDGE AVENUE TROY, OHIO FORM (4-95)

S420 Barbecue Assembly Manual

Blue Series Hot Melt Hose with RediFlext II Hanger System

Instructions for Converting Range to Operate on Liquefied Petroleum Gas

S405 Barbecue Assembly Manual

HOTEL SERVICE HOUSEKEEPING / FOOD + BEVERAGES. ZP Series. A first class service for guests who deserve the best

Installation Instructions

Heyinovo. Disposable Biopsy Valves User Manual. Technical Publications. Document No: WI-RD-13-M07, Rev. B

Legacy Basic Acoustical Shell

PM-X IR SURGICAL POWER & ACCESSORIES

13401 Brooks Drive Baldwin Park, CA p f shademakerusa.com

PEAK 1 GENERAL INFORMATION 1.1 THE CONCEPT 1.2 SAFETY 1.3 THE SPEED SYSTEM 2 THE EMERGENCY PARACHUTE 2.1 EMERGENCY PARACHUTE ASSEMBLY

Hose Reel Series L701/G701

Nova. Rollator OWNERS HANDBOOK. Product Code: WA01063

Swagelok Thermoplastic Hose Hand Swager User s Manual

STERILIZATION and PREPARATION

Meat Slicer INSTRUCTION MANUAL CAUTION! ONE YEAR LIMITED WARRANTY

Bi Wave Plus. User Guide. LFT4906 Issue 5

Operator's Manual. Model: RY10MK-PRO MPN: RA-MLT Gallon Direct Fire Melter Kettle Burner Model with Adjustable Flame-Out Valve

Capacity kw. Thermal turndown from specified maximum capacity 8:1 8:1

User Guide. Tripod. flowtech 75 Tripod. Part No. V

Thermostatic Griddle Field Service Kit Instructions

1000M MANUAL Signature Series Slicer OPERATORS MANUAL

LDR Brachytherapy Seed Sterilization and Sorting Tray

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

Bi Wave Carer. User Guide. LFT5104 Issue 5

Glass Door Fittings Glass Door Accessories

Installation Instructions

INSTALLATION INSTRUCTIONS AND OWNER'S MANUAL AWNINGS. For trailers & motor homes with straight sides TOOLS REQUIRED

Est A TRADITION OF EXCELLENCE INSTRUCTION BOOK.

Rhino-Rack FOXWING ECO Awning 31117

Installation and operation instructions for KINEMAX burners

Advantage Plus TIM-3600 Series Reel

Anchor Family Product Manual for the Rifton Anchors

TUCSON PATIO AWNING INSTALLATION MANUAL

AUTO REWIND HOSE REEL WITH 50 FOOT HOSE. Model 46320

Rhino-Rack FOXWING Awning by OZTENT

OPERATOR INSTRUCTION MANUAL INCLUDING REPAIR PARTS FOR MODULAR GENERAL PURPOSE TENT SYSTEM (MGPTS) TYPE I

SERVICE LETTER. Conquest. WARNING: Further flight with a cracked carry through spar cap is prohibited. The carry

JARVIS. Model 70 Airsnip Air Powered Scissors

The tried-and-tested complete range

INSTRUCTION SHEET DETHATCHER KIT. STS-422LX LPSTS-42JD Assembly Installation Operation Repair Parts L-1660-D. Visit us on the web!

LP/High Altitude LP/High Altitude Natural Gas Conversion Kit For United States Installations

Therme storage water heater. Table of contents. Symbols used. Symbols used Symbol indicates a possible hazard. Installation instructions

LBH18BPK. Instruction Sheet

Installation and Maintenance Manual

NORTHFIELD CORPORATION 1870 COMMERCE DR. DE PERE, WI UNITED STATES

AND LOAD CANOPY RACK SPECIFICATIONS

Thanks for shopping with Improvements! 10 Solar Lighted Gazebo Item #531753

LM601 Landmark Permanent Blind

Electric 2-STAGE SHARPENING FEATURES: system allows for safe, easy knife sharpening - Non-slip feet allows sharpener to stay in place while in use

Butler Personnel Parachutes

Installation & User Manual EW5000 Outdoor Cooking Fire

TUBULAR BURNER CONVERSION KIT AGCK-TBXX / AECK-UBXX

INSTALLER AND OWNER GUIDE

SERVICE LETTER. This service letter applies to the models and serials that have 12,000 or more flight hours as follows

Transcription:

The ULM CASE III Emergency Case Description and instructions for use

Contents 1. Overview.............................. 3 1.1 Special symbols on the device............ 4 2. Description............................. 5 2.1 Purpose............................. 5 2.2 Function............................ 5 3. Safety Instructions....................... 5 4. Use................................... 7 4.1 Ampoule Tray........................ 7 4.2 Dividers............................. 9 4.3 Tube holder.......................... 9 4.4 Ampoule Strips....................... 9 4.5 Ampoule Panels...................... 10 4.6 Cylinder Clamp Straps................. 10 5. Hygienic preparation.................... 11 6. Functional Control...................... 11 7. Maintenance.......................... 12 7.1 Intervals............................ 12 7.2 Disposal............................ 12 8. Extent of Supply....................... 13 8.1 Standard supply schedule.............. 13 8.2 Accessories and Spare Parts............. 13 9. Technical Data......................... 16 10. Warranty............................. 17 11. Declaration of conformity................ 17 2 EN

1. Overview 1 2 27 3 26 25 4 28 24 23 22 21 20 5 6 7 29 8 19 a b 9 10 11 12 30 18 17 16 15 14 13 Legend 1 Cover plate 2 Lock 3 Holding device 4 Lock 5 Top of case 6 Top recess 7 Short top divider (105 mm) 8 Retaining clamps 9 Short base divider (121 mm) 10 Oxygen cylinder recess 11 Bottom recess 12 Base of case 13 Corner of case 14 Case fastener 15 Case handle 16 Long base divider (211 mm) 17 Seal 18 Medium base divider (147 mm) 19 Tube holder a) Oropharyngeal tube, small b) Oropharyngeal tube, large 20 Case hinge 21 Plug-in fastener 22 Rubber tie 23 Long top divider (246.5 mm) 24 Container rack 25 Socket 26 Spring steel sheets Accessories 27 Set of 2 clamp straps for oxygen cylinders 28 Laryngoscope pouch 29 Ampoule strip 30 Ampoule panel Overview EN 3

1.1 Special symbols on the device 2 1 No. Symbol Meaning ULM CASE III device information plate SN Serial number of device 1 Year of manufacture CE marking Other markings 2 Removing the bars 4 EN Overview

2. Description 2.1 Purpose The ULM CASE III Emergency Medical Case is designed for storage and transport of emergency medical apparatus, instruments, drugs and other items. 2.2 Function The interior of the ULM CASE III can be subdivided to suit individual requirements using the variable dividers. Accessories, such as ampoule strips, ampoule panels, tube holders and oxygen cylinder clamp straps, offer further possibilities for individual outfitting. 3. Safety Instructions For your and your patient's safety as well as in accordance with offi cial Safety Regulations (Directive 93/42 EEC) we may point out the following: Please observe the instructions in the User s Manual. The manual is a part of the case and must be kept available at all times. Thorough knowledge of and compliance with these operating instructions are essential when using the case. Description EN 5

The case is intended only for the designated purpose described. If third-party items are used, functional failures may occur and fitness for use may be restricted. Biocompatibility requirements may also not be met. Please note that in such cases, any claim under warranty and liability will be voided if neither the accessories nor genuine replacement parts recommended in the instructions for use are used. Make sure that any leakage of liquids, e.g., from ampoules, is cleared up immediately. Leaked liquids can damage the case and injure the user. Observe the relevant manufacturer s warnings and disposal instructions. To prevent infection and bacterial contamination comply with section 5. Hygienic preparation on page 11. If the case is equipped with oxygen cylinders or other devices, be sure to observe the relevant User s Manuals and special instructions for those devices. Keep the device free from oil and grease. Be sure to wash your hands before changing cylinders. When fi tting or changing cylinders, tighten all screwed connections by hand only. Never use tools for this purpose. Always open the cylinder valve slowly. Smoking and open fl ames are strictly prohibited in the vicinity of equipment containing oxygen. Always ensure that oxygen cylinders contain an adequate supply. It is advisable to keep a full oxygen cylinder in reserve. 6 EN Safety Instructions

4. Use 4.1 Ampoule Tray After opening the case, loosen the rubber ties by lateral pressing the plug-in fastener. Now, hold the spring steel sheets and swing out the ampoule tray. To open the cover plate, move the fastener and turn up the cover plate until the holding device snaps in. Use EN 7

The open cover plate is fi xed. To close it, release the holding device and close the cover plate until the fastener snaps in. The ampoule tray can be removed from the container rack. For this, press the spring steel sheets until the seal is released and remove the tray. To reset the ampoule tray, ead the pins into the corresponding openings in the container rack and press down the tray until the lateral fasteners snap into the spring steel sheets. 8 EN Use

4.2 Dividers The case can be divided internally as required using the dividers. An example is shown on the folding page. When fitting the dividers, please ensure that: they are positioned correctly as in figure they are fitted perpendicularly into opposing grooves the spring fittings are properly engaged To remove a divider, press the two circular spring fittings inwards and lift the divider upwards. 4.3 Tube holder When the case is supplied, the tube holders are fixed to the base, medium divider 18. The tube holder can be moved to another divider in the base of the case by loosening the fixings and relocating the holder as required. 4.4 Ampoule Strips The position of the ampoule strips on the ampoule tray can be changed by loosening the fixings and relocating the ampoule strips as desived. Further strips, e.g. WM 15140 29 available as accessories, can be fixed on the tray as desired. Use EN 9

4.5 Ampoule Panels The ampoule panels WM 8771 30, available as accessories, can be fitted as required in the top of the case simply by fitting the case top dividers in the appropriate position. The position of the ampoule strips on the ampoule panels can also be changed by loosening the fixings and relocating the ampoule strips as desired. 4.6 Cylinder Clamp Straps To fit 0.8 l or 2.0 l oxygen cylinders in the base of the case, use the accessory WM 15141 set of 2 clamp straps for oxygen cylinders 27. The cylinders are fitted in the oxygen cylinder recess as shown. To secure a 2.0 l cylinder, use the outer clamps. The centre and left-hand clamps are used for an 0.8 l cylinder. Ensure that the straps are tight to prevent the cylinder from slipping during transport. 10 EN Use

5. Hygienic preparation Clean the case inside and out simply by wiping with suitable disinfectant solution commercially available. We recommend terralin protect (following the Instructions for Use!). You are recommended to wear suitable gloves (e.g. household or disposable gloves) during disinfection procedures. Remove the dividers first. If the ULM CASE III is fitted with oxygen cylinders, use a clean, dry cloth or a cloth dampened with clean water to clean the outside of the fittings. 6. Functional Control Carry out a functional control after every use(or at least every 2 years if not used). If you discover faults during the functional control, you must not use the emergency case again until the faults have been rectified. Carry out the functional control as follows: 1. Check whether the inserts,cover plate, ampoule strips, corners of the case and the case handle are free from damage. 2. Check that the case fasteners can be opened and closed easily. 3. Check the case and cover plate hinges. They must move easily and be fixed in place correctly. Replace faulty parts. Hygienic preparation EN 11

7. Maintenance 7.1 Intervals Emergency cases are maintenance-free. The deadlines for the regular functional control must be complied with,however (see 6. Functional Control on page 11). We recommend that maintenance work such as inspections and repairs be performed only by the manufacturer, WEINMANN Emergency, or by qualified technicians expressly authorized by WEINMANN Emergency. 7.2 Disposal Do not dispose of the unit with domestic waste. To dispose of the unit properly, please contact a licensed, certified electronic scrap disposal merchant. This address is available from your Environment Officer or from your local authority. The unit packaging (cardboard box and inserts) can be disposed of as waste paper. 12 EN Maintenance

8. Extent of Supply 8.1 Standard supply schedule ULM CASE III WM 8830 Case, fully assembled with: 1x Ampoule tray (incl. 2 ampoule strips) WM 8803 1x Base divider, long (211 mm) WM 8741 1x Base divider, medium, with 2 tube holders (147 mm) WM 8739 1x Base divider, short (121 mm) WM 8743 2x Top dividers, long (246,5 mm) WM 8744 1x Top divider, short (105 mm) WM 8745 8.2 Accessories and Spare Parts Standard symbol labels: Oxygen WM 0900 Put on protective gloves WM 0970 Dressing box WM 0971 Syringes WM 1040 ECG WM 1041 Intubation WM 1042 Infusion WM 1043 CIRCULATION drugs WM 1044 RESPIRATION drugs WM 1045 Universal drugs WM 1046 Resuscitation WM 1047 Baby resuscitation WM 1048 Cannula WM 1070 Extent of Supply EN 13

Tubes WM 1071 Clamps WM 1072 Poison WM 1073 Infant, red WM 1074 Infant, green WM 1075 Wall mount WM 5605 Base divider, long (211 mm) WM 8741 Base divider, medium (147 mm) WM 8742 Base divider, short (121 mm) WM 8743 Top divider, long (246,5 mm) WM 8744 Top divider, short (105 mm) WM 8745 Ampoule panel WM 8771 Laryngoscope pouch WM 8775 Ampoule tray (incl. 2 ampoule strips) WM 8803 Ampoule panel without ampoule strips WM 15139 Ampoule strip* WM 15140 Set of 2 oxygen cylinder clamp straps WM 15141 Holder for oropharyngeal tubes, small* (for infants and neonates) WM 15142 Holder for oropharyngeal tubes, large* (for adults, juveniles and children) WM 15143 Lock WM 15144 Case handle* WM 15145 Seal, set of 2* WM 15147 Ampoule strip, set of 3 WM 15157 Spring steel sheets, set of 2* WM 15162 * including fixings To replace the following fittings, the rivets of the parts needing replaced have to be bored out. Blank rivert tongs will be required. 14 EN Extent of Supply

Case support, set of 2* WM 15149 Case hinge, set of 3* WM 15150 Case fastener, set of 2* WM 15151 Retaining clamp, set of 6* WM 15152 Case corner, set of 4* WM 15158 Cover plate* WM 15159 Holding sheet* WM 15160 Rubber tie* WM 15161 To change the following inserts, the entire case shell has to be dismantled. This involves boring out the rivets from all fittings such as fasteners, hinges, etc. and fitting new rivets after changing the inserts. Blank rivet tongs will be required. Recess, top, variable* WM 15153 Recess, bottom, variable* WM 15154 * including fixings Pull seals available from www.esto.de Pull Tight pull seal 250 mm red, sealing tape Ø 2.3 mm Pull Fly pull seal 125 mm, sealing tape Ø 1.8 mm Bi-Lok pull seal 215 mm, sealing tape Ø 1.9 mm Extent of Supply EN 15

9. Technical Data ULM CASE III Product class as per 93/42/EEC I Standards met Dimensions (WxHxD*) Weight, empty Temperature range: Operation Storage Material: Case shell Case insert Dividers Cover plate E DIN 13232 EN 1789 526 x 400 x 230 8.3 kg 20 C to +50 C 30 C to +70 C anodised Al ABS ABS PC *Outside dimensions incl. corners, supports, handle, etc. Subject to technical change without notice 16 EN Technical Data

10. Warranty WEINMANN Emergency gives the customer a limited manufacturer warranty on new original WEINMANN Emergency products and any replacement part fitted by WEINMANN Emergency in accordance with the warranty conditions applicable to the product in question and in accordance with the warranty periods from date of purchase as listed below. The warranty conditions can be downloaded from www.weinmann-emergency.com on the Internet. We can also send you the warranty conditions on request. In the event of a claim under warranty, contact your specialist dealer. Product WEINMANN Emergency devices including accessories (except masks) for oxygen medicine and emergency medicine Warranty period 2 years Masks including accessories, rechargeable batteries, batteries (unless quoted differently in the technical documentation), sensors, tube systems 6 months Disposable products None 11. Declaration of conformity WEINMANN Emergency Medical Technology GmbH + Co. KG declares herewith that the product complies fully with the respective regulations of the Medical Device Directive 93/42/EEC. The unabridged text of the Declaration of Conformity can be found on our website at www.weinmann-emergency.com Warranty EN 17

Manufacturer WEINMANN Emergency Medical Technology GmbH + Co. KG Frohbösestraße 12 22525 Hamburg GERMANY T: +49 40 88 18 96-120 E: customerservice@weinmann-emt.de Center for Production, Logistics, Service WEINMANN Emergency Medical Technology GmbH + Co. KG Siebenstücken 14 24558 Henstedt-Ulzburg GERMANY WM 16130g 04/2017 EN Simply Professional Made in Germany WEINMANN-Emergency.com