RM 2125 Troubleshooting Guide

Similar documents
TN5000 Titan Manual Rotary Microtome

Instructions for Cryostat Slicing Leica Model CM 3050S

MODEL # 1 MANUAL CAN OPENER

Leica CE/ Leica CN. Knife holder

SERVICE MANUAL- M126

Sectioning Paraffin Blocks

Leica RM2125 RTS. Rotary Microtome. Instructions for Use

Product instruction manual Ream Cutting Systems RE3943, RE3946, RE3947, RE3971, RE3952E

M129 SERVICE MANUAL- M129 MODEL #G-2 MANUAL CAN OPENER

JARVIS. Model 70 Airsnip Air Powered Scissors

DIAMOND TECH INTERNATIONAL Innovations For Creativity DIAMOND LASER BAND SAW OPERATIONS MANUAL

RM2235. Rotary Microtome. Instructions for Use

Maintenance and care of the. German Red Clipper and its blades

1000M MANUAL Signature Series Slicer OPERATORS MANUAL

MAX Series 9514 OPERATOR S MANUAL

WARNING: There is a possibility of death, serious injury, or physical loss resulting from improper use by ignoring this indication.

functionality with the necessary maintenance tools stored in a convenient built in tool box, as well as parallel handles for

Est A TRADITION OF EXCELLENCE INSTRUCTION BOOK.

PRODUCT MANUAL - M096

GLM300 Semi-auto Slicer

PS-10 PS-12D PS-12. Primo Slicer Manual FOR OPERATOR-DO NOT DISCARD

Professional Mandoline Manual. Model: 90797

4A1-114/114KR & 4A1-2 MANUAL TENSIONERS

Notes, Parts List & Drawings

ROTARY MICROTOME MODEL MRP 2015 PROCEEDING MANUAL

Downtime equals wasted time. Shandon Finesse Microtome Series

Installation Instructions for the Rolltec Bravo Awning

Maintenance Instructions and Directions for Use Wind Support

Mixer Blade Sets of Parts and Installation Procedure Using Epoxy

S E L E C T I O N. Abdominal Crunch. User manual

1612P MODEL 1612P SLICER MODEL EXECUTIVE OFFICES 701 RIDGE AVENUE TROY, OHIO FORM (4-95)

TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM INSTRUCTIONS

INSTRUCTION SHEET TWISTLOCK VEHICLE STABILIZER (TVS) SYSTEM PN S: , ,

INSTALLATION & MAINTENANCE MANUAL SERIES 9000 MODELS 9150/9300 SERIES 9000 VALVES WITH V BALL

Thomas Scientific Swedesboro, NJ U.S.A.

300 Model 870 Four Wheel Hose Cart Owner s Manual

Application Tooling Specification

IMPORTANT NOTICE. Professional Mandoline Manual

AND LOAD CANOPY RACK SPECIFICATIONS

Operating Instructions

Professional Mandoline Manual. Model: 90757

2008 Equalizer Industries, Inc. Revision 4/23/

42" SNOW BLADE. Owner's Manual. Model No Safety Assembly Operation Maintenance Parts

PATRIOT INSTRUCTION MANUAL SL300C 12 MEAT SLICER 1/2 HP

Wenger Corporation 2014 Printed in USA 05/14 Part #049B015-09

SEMI AUTOMATIC PARAFFIN MICROTOME MODEL MRP 2016 SA PROCEEDING MANUAL

INSTALLATION INSTRUCTIONS for folding arm awnings

CHANGING YOUR LANDSCAPE SINCE 1945 OWNER S MANUAL. Tow Hitch Replacement Kit For Rough Cut Trailcutters. Starting Serial # L

8713S Premium Slicer Operators Manual

Important! Read all of these instructions before assembling or riding the glider. For questions or help please call Glide Bikes at

Tidland Series 'C' Knifeholder MAINTENANCE MANUAL Crush Cartridge Class I, II, III Crush Slitting

Owner/Operator Manual

PIONEER LITE PATIO AWNING OWNER'S MANUAL

Thermo Scientific Sliding Microtome Microm HM 450 INSTRUCTION MANUAL. Thermo Scientific Sliding Microtome Microm HM 450

Tidland Class II Razor Cartridge

User s manual EFS-150 USER S MANUAL EFS-150

Installation and User s Manual 12 x 10 MOTORIZED AWNING

IMPORTANT OWNER-OPERATOR INSTALLATION INSTRUCTIONS. Part # A7006

I N ST R UC T I ON MODELS 1612E, 1712E & 1712RE SLICERS. FORM Rev. B (10-97) 1712RE 701 S. RIDGE AVENUE TROY, OHIO 45374

CRD120SC TUBING CUTTER

Contents. Awnings USA - Full Protective Hood Manual Instructions ft 11" - 11ft 6" Awnings

Swagelok Thermoplastic Hose Hand Swager User s Manual

NEWMAR SERVICE SCHOOL

SECTIONING PROCEDURE FOR LASER CAPTURE MICRODISSECTION

Solera Universal Hardware

STOP BOX SCRAPER/GRADER BLADE. Owner's Manual. Model No Safety Assembly Operation Maintenance Parts

Tuke Saw System. Operating Manual. Safe Navigation Through Soft Tissue

Urea/Adblue Hose Reel

Synflex Assembly Equipment and Accessories

Two Piece Low Torque Kelly Valve

American Classic Rear Terrain Hub Bearing Change, Cassette Body Swap and Overhaul Instructions.

100% MADE IN THE USA TRIPOD AND MONOPOD MANUAL

FORM (Apr. 2006)

GENERATION III COULTER SET-UP & PARTS MANUAL

1000S Premium Automatic Slicer Operator s Manual

OPERATIONAL & MAINTENANCE MANUAL SEM Model 1400 Series Disintegrator Rev. 2

Installation Instructions for the Rolltec Adalia X3M Extenda Awning

How to operate TBC-553L

INSTRUCTION MANUAL HYDM140

Installation Instructions for the Rolltec Physique XL Awning

Knife Permit: Outdoor Knife Skills

The Groomers Scissor Guide

Meat Slicer INSTRUCTION MANUAL CAUTION! ONE YEAR LIMITED WARRANTY

Tools. For safety and easy faucet replacement, San-Ei recommends the use of the following helpful tools:

accidents which arise due to non-observance of these instructions and the safety information herein. CAUTION:

LOAD BINDER WINCHES AUTO TENSIONER (AIR WINCH) PART NO 30/303899

Tuf-Lite and Tuf-Lite II Fans 6000M Series Hub

Model 205 Fireview Maintenance Kit

Mandolin Slicer Quite possibly the safest mandolin in the world

Instruction Manual for Globe Slicer Models 3600P, 3850P, and 3975P

Instructions for Use A-dec Quick-Disconnect Syringes

INSTALLATION INSTRUCTIONS

Auto Rewind Hose Reel 10M / 20M / 30M Models

THE EDGE12 SLICER. MODEL EDGE12-2 (220/50/1 voltage) FORM (July 2012) EDGE12 SLICER

Instruction Manual for Globe Slicer Models 3600, 3850, and 3975

ALEKO Victoria Series Retractable Awning Instructions for Installation and Operation

Setup. Assembling. Attaching the sharpening stone

USER MANUAL BMU WORKCAGE. CONFORM TO THE MACHINE DIRECTIVE 2006/42/EC and to EN1808 (1999)

FOR CHEFMATE SLICER MODEL GC510

DeZURIK AM-SERIES MANUAL GEAR ACTUATOR USED ON BUTTERFLY VALVES

Transcription:

Problems, possible causes and corrective action Problem Cause Corrective Action 1) Thick and Thin Sections. a) Partial Sections b) Shorter Sections in ribbon c) Section thickness varies from one section to another d) Skipping sections e) Varied thickness in one individual section RM 2125 Troubleshooting Guide I) The clearance angle is too small. a) Systematically try several clearance angle adjustments until the optimum angle is found. II) Insufficient clamping of specimen. -- Orientation head -- Cassette clamp b) Check if all levers are locked and screws are tightened on the specimen. Retighten the levers and screws, if necessary. c) If the clamp lever can be tightened at more than a 60 degree turn, adjustment is needed. d) Open up a service call for the adjustment of the specimen head clamp lever. e) Check 4 screws that hold orientation head to the advance arm. f) On the cassette clamp, check the 4 screws that hold cassette clamp to the orientation head. Status: May 2003/SlRocco/edited WE 1

Problem Cause Corrective Action g) Check for movement of cassette while in holder. If movement exists, it may be caused by: -- Weak/Broken Springs -- Wax buildup -- Lack of grease A service call is recommended to replace springs, and properly grease. If it is wax build up, do clean with a sparing amount of Xylene or chemicals to remove paraffin. III) Locking levers are not well lubricated or cleaned, causing them not to be completely fastened or stuck in a non-locked position. h) Remove all locking levers, clean off all paraffin and debris, then apply a sparing amount of the provided Leica lubricant. Lubricate only with grease recommended by the manufacturer. IV) Leaf spring worn. i) Replace leaf spring. Call your nearest Leica Microsystems dealer for order information. j) If the leaf spring does not solve the problem, then replace the locking lever. A Service call is required for this. V) Dull knife/blade. k) Use a different part of the cutting edge or use a new knife/blade. VI) Rotation in specimen advance system. l) Try to rotate the object head clockwise and counter clockwise. If there is play in the system, open up a service call for the internal adjustments of the specimen advance system. 2) Compressed sections. The sections are extremely compressed, wrinkled or jammed together. a) Compressed b) Wrinkled c) Jammed I) Dull knife/blade. a) Use a different part of the cutting edge or use a new knife/blade. Status: May 2003/SlRocco/edited WE 2

Problem Cause Corrective Action II) Excess of oil on blade III) Blade/knife is gummed up with paraffin b) Remove oil with paper towel or Q-tip. Be aware of sharp blade edge. c) Clean knife with a swab moistened IV) Specimen too warm. d) Pre-cool the specimen on a cold plate. e) Cool the specimen in iced water or with an ice-cube. f) Use freeze spray or ice cube immediately before sectioning. V) Clearance angle too big. g) Clearance angle adjustment; systematically decrease the clearance angle until the optimum adjustment is 0 obtained. h) Check leaf spring screw for washers and tightness. VI) Leaf spring worn i) Replace leaf spring. Call your nearest Leica Microsystems dealer for order information. j) If the new leaf spring does not solve the problem, then replace the locking lever. A Service call is required for this. k) Check angle of pressure plate facet to see if correct. Refer to User Manual. Status: May 2003/SlRocco/edited WE 3

Problem Cause Corrective Action VI) Paraffin buildup on the back of the pressure plate. k) Clean knife with a swab moistened VII) Knife blade not parallel with pressure plate. Blade Pressure Plate l) Adjust Pressure Plate with set screws at bottom of knife holder (turn counter clockwise or clockwise) until blade is parallel with the top of the pressure plate. Blade Pressure Plate Blade Edge m) When the blade is parallel with the pressure plate, the base line of the blade should line up with the top of the pressure plate after being clamped. Pressure Plate Base line of blade VIII) Sectioning speed too high. n) Rotate the hand wheel at a slower speed. *NOTE) If hand wheel is too loose and slowing sectioning speed is not possible, open service call for replacement of tensioning washers and make proper adjustments to hand wheel torques. 3) Chatter a) Venetian Blind effect b) Washboard c) Undulations I) Sectioning speed too high. a) Rotate the hand wheel at a slower speed. II) Clearance angle too big. 0 b) Clearance angle adjustment; systematically decrease the clearance angle until the optimum adjustment is obtained. Status: May 2003/SlRocco/edited WE 4

Problem Possible Cause Corrective Action III) Damage to pressure plate exists: c) Replace pressure plate. -- Bent -- Nicked Call your nearest Leica Microsystems dealer for order information. IV) Insufficient clamping of specimen and/or knife. d) Check if all levers are locked and screws are tightened on the specimen knife holder systems. Retighten the levers and screws if necessary. V) Incompatible blades. Blade is possibly too thin causing flex if the specimen is too hard. VI) High-profile knife may be too thick. VII) Wax build-up on the back of the front pressure plate. e) Try using different brand and grade blades. f) Try using a different brand of blades. g) Clean back of pressure plate. VIII) Application h) -- Infiltration -- Dehydration -- Decalcification -- Paraffin is too old -- Paraffin Impurities IX) Tabletop stability. X) Vibration in specimen advance system. I) Move the microtome or table to a more stable location, or replace table if table is flimsy j) Check if orientation has been changed or disassembled incorrectly. Re-install all parts if necessary. 4) Scratches or splits in sections I) Knife: -- Defect in knife edge -- Hard specimen in block -- Paraffin on front or back of knife edge a) Move to a new area of the knife and observe if scratches move with it. If so, replace knife or have knife sharpened. b) Notify supervisor about hard specimen in block and how to proceed from there. Status: May 2003/SlRocco/edited WE 5

Problem Possible Cause Corrective Action c) Clean knife with a swab moistened d) Try to move blade or replace. II) Blades: -- Defect in knife edge -- Hard specimen in block -- Paraffin on front or back of knife edge -- Coating deterioration on blade edge TIP) Use a wooden stick and run it across the blade once and check again. e) Clean pressure plate with a swab moistened in Xylene. a) Move to a new area of the knife and observe if scratches move with it. If so, replace knife or have knife sharpened. b) Notify supervisor about hard specimen in block and how to proceed from there. c) Clean knife with a swab moistened d) Try to move blade or replace. e) Remove paraffin from front or back of the pressure plate. f) Clean pressure plate with a swab moistened in Xylene. Status: May 2003/SlRocco/edited WE 6

4) No specimen advance and consequently no section produced. Instrument errors, possible causes and corrective action Problem Possible Cause Corrective Action I) Specimen reached the front feed limit. a) Turn the coarse feed wheel in the appropriate direction to move the specimen towards the rear limit. II) User may accidentally be holding coarse feed wheel while sectioning. III) The coarse feed wheel cannot rotate freely. IV) Internal mechanism (gear) may be defective. b) Remove any obstruction. c) Open service call for replacement of course advance mechanism. 5) Drifting of hand wheel. I) Hand wheel is not balanced properly. Handwheel should remain at 12 o'clock when positioned at that location. No drifting should occur. a) Adjustment of hand wheel position and/or collar sleeve position. Service call is needed. 6) Specimen is picked up in the return stroke of the specimen arm. 7) Sections sticking to pressure plate. I) Static electricity charge may be built up on the knife holder or specimen head. a) Adjust temperature or humidity level in the room. b) Use an ionizer to neutralize the static charge. c) Clean components of the microtome with alcohol. I) Too much paraffin. a) Clean knife with a swab moistened II) Dirty pressure plate. b) Clean plate with a swab moistened with Xylene. Status: May 2003/SlRocco/edited WE 7

Problem Possible Cause Corrective Action III) Static c) -- Spray static solution -- Increase humidity -- Ground the instrument IV) Too much Xylene on pressure plate. d) Remove Xylene with alcohol. 8) Unable to retract specimen NOTE) This is only on the RM 2125 RT model. I) Retraction mechanism on specimen advance system is defective. a) Open a service call. 9) Microtome makes "clunking noise" when sectioning. I) Linear bearings in cage are at their lower limits. a) Linear bearings need to be cleaned and possibly re-greased. II) Alignment of cage is incorrect. b) Open a service call. Status: May 2003/SlRocco/edited WE 8